© Adam Equipment Company 2010
•
When a parameter is entered by pressing the
[Tare]
key, the displays
will guide you through the parameter selected and the options
available.
The parameters available are:
“F1”
“
Unit”
Default Weighing Unit
“F2”
“
dPP”
Decimal Point Position
“F3”
“
CAP”
Select capacity and increment
“F4”
“
CAL”
To enter the Calibration
“F5”
“
init”
“ZEro”
Initial Zero Range
“F6”
“
rE”
“ Zero”
Re-Zero range
“F7”
“
SCSivE” “ tArE”
Successive Tare Enable
“F8”
“
Ad”
“CoUnt”
Display the A/D counts
“F9”
“
Pin”
To set the password
“F10”
“
Lvd”
“mode”
Low voltage detection
“F11” “trACk” “ZEro”
Zero track setting
12.2.1
F1 –DEFAULT WEIGHING UNIT
To set the weighing unit that will be used initially for calibration and setting the
calibration. This weighing unit will always be available to the user.
•
To enter this parameter, press the
[Tare]
key when
“F1”
is shown. The
display will show the current setting.
•
Press the
[U.Wt./Units]
key to change the value. Select from the units
displayed, typically kilograms (kg), grams (g) or pounds (lb) for USA
sales and kilogram or grams for the rest of the world.
•
Press
[Tare]
to accept the displayed value.