© Adam Equipment Company 2010
7.0 USER
PARAMETERS
The parameters are set to customise the scale to suit the weighing applications. You
need to enter a secure menu by entering a password when requested.
•
Press
[Tare]
once, during the initial counting of the display after the
power is turned on.
•
The
“Weight”
display will show
"PIn "
requesting for the password
number.
•
The default password is
"0000"
but other numbers can be set using the
parameter menus. Press the
[0]
key four times.
•
Press the
[Tare]
key.
•
There are 3 functions that can be cycled through using the
[U.
Wt./Units]
key. The
Weight window
will show the parameter number
and the
Unit Price window
will show the word describing the function.
•
The first parameter is
“F1” “CAL”
. (See section 8.0 for details)
•
To enter into Calibration function, press the
[Tare]
key.
•
To exit a parameter, press the
[Zero]
key.
•
To select the second parameter press the
[U. Wt./Units]
key.