12.2.2
F2 –DECIMAL POINT POSITION
•
To set the value for the decimal point location. The options are 0, 0.0,
0.00, 0.000, 0.0000
•
To enter this parameter, press the
[Tare]
key when
“F2”
is shown.
•
The display will show the current setting.
•
Press the
[U.Wt./Units]
key to change the value. Select from 0, 0.0,
0.00, 0.000, 0.0000
•
Press
[Tare]
to accept the displayed value.
12.2.3
F3 – CAPACITY
•
To enter this parameter, press the
[Tare]
key when
“F3”
is shown. The
display will show the current capacity.
•
Enter the numeric values using the keypad. The scale will check the
number of divisions n = maximum/increment is less than 30,000
divisions.
•
Press
[Tare]
to accept the displayed value.
NOTE:
The scale use load cells suitable for the capacity selected.
•
The display then will go to the increment setting.
•
To set the value for the increment of the scale. For example 100kg x
0.01kg the increment is 10grams. but the last digit increments by 1.
•
The display will show the current increment value.
•
Press the
[U.Wt./Units]
key to change the value. The indicator will
determine what the minimum and maximum increment is possible for
the capacity and decimal point you have set.
•
Press
[Tare]
to accept the displayed value.