Predefined linearization curves
•
Linear - no linearization
Volume linearization
•
Horizontal cylindrical vessel
•
Spherical vessel
Flow linearization
•
Venturi, trapezodial weir, rectangle weir
•
Palmer-Bowlus-Flume
•
V-Notch, triangle weir
Free linearization characteristic with up to 40 points
•
Custom
Default > Linear
Input mode custom defined linearization
•
Percent / Linearized Percent
•
Filling level / Display Scaling
•
CSV-file via Bluetooth-interface
The percentual input (referring to the measuring range) is only possible without applying measuring
value, whereby the input in filling level and display scaling e.g. liter can be done without applying
measuring signal as well as with applying measuring signal.
At the linearization without applying measuring signal, for each linearization point a needed signal
value (in percent or filling level) is input and referred to the needed output value (in percent or
primary unit) that must be also input.
At the linearization with applying measuring signal, for each linearization point the current measuring
value is captured and referred to the output value (acc. to display scaling) that must be input.
•
Linearization Points
The number of linearization points for the complete measuring characteristic must be defined.
The input range is limited to values from 2 to 40.
Default > 2
•
Linearization
At the input mode percent / linearized percent the input signal refers percentual to the
measuring range, that is defined at the min./max. adjustment for 0% and 100%. The output
signal must also be considered percentual.
At the input mode filling level / display scaling the input signal refers to the filling level
(inverted distance). The output signal refers to the display range, that is defined in the display
scaling.
The complete characteristic, eventually also the measuring range end values 0% and 100%
must be defined, cause the measuring range end values from the min./max. adjustment are
not copied into the linearization table.
The input range is indefinite.
Default > Linearization point 1 > 0.000% = 0.000% resp. 0.000bar / Linearization point 2 >
100.000% = 100.000% resp. 1.000bar
•
Store
Input linearization points are not automatically stored loss-protected.
To store one or also more linearization points loss-protected, the function store must be
executed.
Sensor TAG
Due to the Sensor TAG different devices can be differentiated. At the indicator type digital the Sensor
TAG is indicated in the display.
The Sensor TAG is added automatically to the Bluetooth-name, to allow the device identification when
using multiple devices in reception range.
Up to 19 characters can be input.
The input range is indefinite.
Default > no Sensor TAG allocated