Factory Reset
The factory reset changes all settings to default values.
The factory reset does not concern:
•
Diagnose data
•
Historical measuring data
•
Storage interval
•
Custom defined linearization
Display
The symbol E at a menu item indicates its position in the extended menu structure.
This extended menu structure can be shown in the submenu display by the menu item menu
structure.
Indicator value
The measuring value can be alternatively displayed as distance, as filling level, percentual, with any
desired scaling referring percentual to the measuring range or by displaying the analogue output
value.
•
Distance
•
Filling level
•
Percent
•
Scaled
•
Signal output
Default > Distance
Scaling
– extended menu structure
This menu item is only available at indicator value scaled.
By defining a scaling the measuring range can be rescaled into any desired numerical range. This
allows e.g. the indication of the volume in liter.
The current measuring value is indicated in the display.
The input of the measuring range limits 0% and 100% are not mandatory. Values within the
measuring range e.g. 11% and 87% can be also input. In this case there is an automatic calculation
to 0% resp. 100%.
•
Lower display value
•
Upper display value
The input range is indefinite.
Default > Measuring value 0.000% = Display 0.000 / Measuring value 100.000% = Display 1.000