46
Display
Causes
Solution
Temperature has exceeded
the operating temperature of
the system
Move to a cooler environment. Repeat
the test.
Temperature is below the
operating temperature of the
system
Move to a warmer environment.
Repeat the test.
Batteries are low
Test results will still be accurate. But
replace the batteries as soon as
possible.
Batteries are discharged and
the meter does not allow more
tests until replacement with
new batteries
Replace the batteries. Repeat the test.
Meter electronics failure
If the problem persists, please contact
Customer Support at 1-800-838-9502.
Non
On Call
®
Express
test
strip inserted in the meter
Please make sure you use the
On
Call
®
Express
test strip with the
On
Call
®
Express Voice
Blood Glucose
Meter.
If the problem persists, please contact
Customer Support at 1-800-838-9502.
Test strip damaged or
calibration error
Please test again by using a new strip.
If the problem persists, please contact
Customer Support at 1-800-838-9502.
Communications failure
There is an error in transferring data to
the PC. See the package insert
included in the PC data transfer
software for troubleshooting. If the
problem persists, please contact
Customer Support at 1-800-838-9502.
Meter data port is plugged in
with the data transfer cable
when meter is in “waiting for
sample application” mode with
strip already inserted into the
meter strip port
Unplug the data transfer cable from
the meter’s data port. Then remove
the strip. Reinsert the strip into the
strip port for testing.
If the problem persists, please contact
Customer Support at 1-800-838-9502.
For additional questions or issues with this product, please contact Customer
Support at 1-800-838-9502, 24 hours a day, 365 days a year.