30
3. Following this display check, the system will enter the test mode. The display
will show the date and time and the strip icon with the blood sample icon
blinking and say “
Please apply blood to the test strip
,” to indicate that the
test strip is inserted correctly and a drop of blood can be added. If the test strip
has been inserted incorrectly, the meter will not turn on. The meter is ready for
testing when the blinking blood drop and strip symbol appears. At this time a
blood drop can be added.
4. Touch the blood sample to the sample tip at the end of the test strip. If the
volume > 0, the meter will beep and voice “
Testing
” to indicate the sample is
sufficient and the measurement has started. If you applied a drop of blood, but
do not see the countdown begin, you may reapply a second drop of blood
within 3 seconds.
DO NOT:
Apply the sample to the front or back of the test strip.
Smear the blood drop onto the test strip.
Press your finger against the test strip.
The meter will count down from 3 to 1. Then the blood glucose level will be
voice broadcast and display on the screen will indicate that measurement is
complete. The measurement’s unit, date, and time of the test will also be
displayed.
READY TO TEST