42
infectious diseases, even after you have performed cleaning and disinfection.
Please follow proper precautions when handling your meter and lancing device.
No signs of meter or lancing device deterioration in appearance or performance
were observed after 260 cycles of the recommended pre-cleaning and disinfection
procedure. However, you should always inspect your meter and lancing device
after performing the cleaning and disinfection procedure. External signs of meter or
lancing device deterioration might include polymer crazing (thin silver streaks),
cracking, swelling, dissolving, softening or becoming brittle. Signs of meter
performance deterioration might include buttons not functioning properly, audio not
working, missing display segments, error messages during testing or control
solution test results that are out of range. Signs of lancing device performance
deterioration might include reduced functionality of the lancing depth adjustment,
arming slider or firing button.
If any signs of external or performance deterioration are observed on your meter or
lancing device after cleaning and disinfection, please stop using your meter and
lancing device and contact Customer Support at 1-800-838-9502 for assistance.