34
Precautions and Limitations
The
On Call
®
Express Voice
meter, the
On Call
®
Express
test strips and other
components have been designed, tested and proven to work together
effectively to provide accurate blood glucose measurements. Do not use
components from other brands.
The
On Call
®
Express
Test Strips test fresh capillary whole blood. Do not use
with serum or plasma samples.
The
On Call
®
Express Voice
Blood Glucose Monitoring System is for self-testing
users to test fresh capillary blood. For over the counter sale. Never test blood that
has been placed in test tubes containing fluoride or other anticoagulants. This
will cause falsely low results.
Very high (above 60%) and very low (below 25%) hematocrit levels can cause
false results. Talk to your doctor to find out your hematocrit level.
The system is tested to accurately read the measurement of glucose in whole
blood within the range of 20-600 mg/dL.
Fatty substances have no major effect on test results. These include
triglycerides up to 3,000 mg/dL or cholesterol up to 500 mg/dL.
Ascorbic acid (vitamin C) when occurring in blood at normal or at high
therapeutic concentration does not significantly affect results. However,
abnormally high concentration (> 3 mg/dL) in blood may cause inaccurately high
results.
The
On Call
®
Express Voice
Blood Glucose Monitoring System has been tested
to work properly up to 8,516ft (2,595 meters).
Not for use on critically ill patients, patients in shock, severely dehydrated
patients or hyperosmolar patients (with or without ketosis).
Dispose of blood samples and materials with care. Treat all blood samples as if
they are infectious materials. Follow your doctor's recommendations for disposal.