background image

Bioland

®

 

Blood Glucose Monitor  

Model

G-427G 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

OWNER'S MANUAL 

 

 
 
 

Summary of Contents for G-427G

Page 1: ...Bioland Blood Glucose Monitor Model G 427G OWNER S MANUAL...

Page 2: ...gned to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in a...

Page 3: ...n liquid nor put it where it could fall into liquid 2 The following basic safety notice should always be taken 3 Use the device only for the intended use described in this manual 4 Do not use adjunct...

Page 4: ...ETER AND DELETING THE MEMORY 8 THE TWO MEASURING MODES 10 TESTING WITH BLOOD SAMPLE 10 Overview of the Lancing Device 10 Setting Up the Lancing Device 11 Preparing the puncture site 12 Performing a Bl...

Page 5: ...e Monitoring System is intended to be used for the quantitative measurement of glucose sugar in fresh capillary whole blood samples drawn from the fingertip The Blood Glucose Monitoring System is inte...

Page 6: ...check that the glucose meter and test strips are working properly and that you are doing the test correctly These solutions contain a known range of glucose as indicated on the strips vial bottles Thi...

Page 7: ...cise or any other event take a blood sample from your finger only You do AST only in the following instances In a pre meal or fasting state more than 2 hours since the last meal Two hours or more afte...

Page 8: ...and simple messages 4 M BUTTON It is used to turn on the meter enter the memory mode 5 S BUTTON It is used to set up the meter Meter Screen Display Message 1 BLOOD DROP SYMBOL Flashes when it is read...

Page 9: ...ts might be wrong if the contact Bar is not fully inserted into the test slot The front side of the test strip should face up when inserting the test strip SETTING THE METER AND DELETING THE MEMORY 1...

Page 10: ...n or OFF press the M button to turn on or turn off the first alarm Press the S button to select On then to set the hour When the hour is flashing press the M button to add an hour When the correct hou...

Page 11: ...TESTING WITH BLOOD SAMPLE CAUTION To reduce the chances of infection Choose a clean dry work surface Never share a lancing device and lancet with another person Always use a new and sterile lancet La...

Page 12: ...lace 2 Screw the needle cap and place it in appropriate location to use again after finish blood taking 3 Replace the lancing device cap Turn the cap until it is snug but not too tight And then Adjust...

Page 13: ...l and let it air dry Fingertip testing Hold the lancing device firmly against the side of your finger Press the release key You will hear a click indicating that the puncture is complete NOTE Choose a...

Page 14: ...hen GSM connected GSM symbol flashing and the test results data will transmission to the Cloud system automatically NOTE The results obtained from this meter are plasma calibrated This helps you and y...

Page 15: ...Fast for at least eight hours and then do comparison tests if possible Take your meter with you to the lab While staying at the lab Make sure that the samples for both tests are taken and tested with...

Page 16: ...the test Expired or contaminated Control Solution Expired or contaminated Test Strip Malfunc on of your Meter If you continue to get control solution test results that fall outside of the range printe...

Page 17: ...the meter works well for example display will be clear to see after cleaning and disinfection can avoid the infection to you or to the other people and the cross infection We strongly warn that the d...

Page 18: ...ycles Operating conditions 50 104 Humidity 25 80 RH 860hPa 1060hPa Storage Condition system meter and test strips 39 104 and below 80 relative humidity 700hPa 1060hPa Storage Condition meter 40 131 Hu...

Page 19: ...low less than 20 or very high more than 60 red blood cell count hematocrit can lead to incorrect test results If you do not know your hemarocrit level please consult your health care provider DISPLAY...

Page 20: ...ent or Control Solution test results are not within the specified range 1 Not enough sample in the Test Strip 2 Test Strip or Control Solution has expired 3 Test Strip has been damaged due to heat or...

Page 21: ...20 19 Version V1 0 2016 01 by Bioland Technology LTD...

Reviews: