8.1
COAXIALITY
INTRODUCTION
The Coaxiality program is used to determine
the direction of the rotational axis of a
secondary spindle or a tool holder, in relation
to a main spindle in a machine tool. The axis
of rotation in the main spindle is used as a
reference and the axis of rotation of the other
object is measured in relation to that. The
result is presented as an angular and offset
deviation in two directions, perpendicular and
parallel to the bed slant.
The Coaxiality program is also used to
monitor the actual position of the secondary
spindle or a tool holder during adjustment.
During adjustment, the program shows the
offset and angular misalignment live in both
directions.
MEASUREMENT METOD
The angular and offset error of the secondary
spindle (or tool holder) is determined by
measuring the distance between the rotational
center of the main and secondary spindle in
two planes.
The MEAX SR sensor is mounted on the main
spindle (reference) and the MEAX SM sensor
is mounted on the secondary spindle (or tool
holder). The distance between the sensors
Summary of Contents for Meax DU 01
Page 1: ...USER MANUAL MEAX...
Page 2: ......
Page 4: ......
Page 8: ...1 4...
Page 19: ...4 3 Do not use paper tissue which can scratch the detector surface Do not use acetone...
Page 20: ...4 4...
Page 24: ...5 4...
Page 56: ...7 12 Save the measurement result Go to adjustment...
Page 60: ...7 16...
Page 76: ...8 16...
Page 79: ...9 3 STARTING THE PROGRAM Start the Squareness program by touching its icon in the Home Menu...
Page 83: ...9 7 Confirm configuration Confirm configuration and continue to summary screen...
Page 104: ...10 10 Go to adjustment...
Page 108: ...10 14...
Page 122: ...11 14...
Page 132: ...12 10 Coaxiality Go to distance Measure and enter distance between the sensors SR and SM...
Page 140: ...12 18...
Page 148: ...13 8 Coaxiality Go to distance Measure and enter distance between the sensors SR and SM...
Page 155: ...13 15 Global Settings Open Global Settings Return Return to the program...
Page 156: ...13 16...
Page 160: ...14 4 Zero values Halve values Reset values to raw data...
Page 161: ...14 5 Sample and display a single value Return to live values...
Page 174: ...17 4 ZERO VALUES Measure max min Reset values to raw data...
Page 187: ...19 9 Level screen Switch to Straightness page...
Page 200: ...19 22...
Page 220: ...22 6...
Page 226: ...23 6...
Page 228: ...24 2...
Page 232: ...25 4...
Page 237: ...26 5 MAGNETIC BASE Pointing fixture and magnetic base...
Page 242: ...27 4...
Page 252: ...30 2...
Page 253: ......