EXC-500 MICROSCOPE SERIES
ACCU-SCOPE
®
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
13
ADJUSTMENT & OPERATION
(continued)
Adjusting the Focus -
(Fig. 19)
To ensure that you obtain sharp images with both eyes (since
eyes vary, especially for those wearing glasses) any eyesight
variation can be corrected in the following manner: set both
diopter collars to “0”. Using your left eye only and the 10X
objective, focus your specimen by adjusting the coarse
adjustment knob
①
. When the image is in view, refine the
image to its sharpest focus by turning the fine adjustment
knob
②
. Rotate the diopter collar to obtain the sharpest
focus. To obtain the same sharp image using your right eye,
do not touch the coarse or fine adjustments. Instead, rotate
the right diopter collar until the sharpest image appears.
Repeat several times to check.
NOTE:
do not counter rotate the focusing knob as this will
cause severe problems and damage to the focusing system.
Adjusting the Focus Stop
-
(Fig. 20)
Once you have adjusted the focus, turn the focus stop
clockwise to set your objective levels to a stop position so they
will not come into contact with your slide.