Initial Startup •
9
139
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
9.10 Value ranges (normal, critical, reportable)
This menu allows you to set limit values for results as
Normal
,
Critical,
or
Reportable
. Results outside these
limits will be flagged to alert you of this event. The
Reportable Range
allows the system administrator to set
an institution defined range for reporting patient results.
Values for range limits are entered one directly after the
other in the following order:
■
Normal (low/high limit value)
■
Critical (low/high limit value)
■
Reportable (low/high limit value)
1
Use the keypad to enter the first value.
2
Touch
to enter the following value,
or
touch
to return to the previous view.
3
After entering the last value, touch
to save all
the previous settings.
1
2
3
4
5
6
7
8
9
0
Report. Range Lo
11
mg/dL
12:48 pm
03/17/17
1
2
3
4
5
6
7
8
9
0
Normal Range Lo
70
mg/dL
12:48 pm
03/17/17
1
2
3
4
5
6
7
8
9
0
Critical Range Lo
40
mg/dL
12:48 pm
03/17/17
1
2
3
4
5
6
7
8
9
0
Report. Range Hi
600
mg/dL
12:48 pm
03/17/17
1
2
3
4
5
6
7
8
9
0
Normal Range Hi
200
mg/dL
12:48 pm
03/17/17
1
2
3
4
5
6
7
8
9
0
Critical Range Hi
301
mg/dL
12:48 pm
03/17/17
Summary of Contents for Inform II
Page 3: ...08424705001 01 2017 04 USA Operator s Manual Version 6 0 A ACCU CCU CHEK HEK Inform II System ...
Page 245: ...245 Notes ...
Page 246: ...246 Notes ...
Page 247: ...247 ...