08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Measurement Flow
Comments required*
0: optional
1: out of range
2: required
3: disabled
0
N
Y
Comments requirement level: if out of…
(only valid if
Comments required
= 1)*
0: normal range
1: critical range
2: reportable range
3: measurement range
0
N
Y
Custom comments*
0: disabled
1: enabled
1
N
Y
* Does not apply to QC measurement workflow
Control lot verification
0: display only
1: yes / no confirmation
2: prompt for entry
3: scan only
1
Y
Y
Control result display
0: value (numeric)
1: PASS / FAIL
0
N
Y
Critical range display
0: value (numeric)
1: HI / LO
0
N
Y
Critical range HI limit
10 – 600 mg/dL
0.6 – 33.3 mmol/L
600 mg/dL
33.3 mmol/L
Y
Y
Critical range LO limit
10 – 600 mg/dL
0.6 – 33.3 mmol/L
10mg/dL
0.6 mmol/L
Y
Y
Critical range message enabled
Whether to display the
out-of-critical range
warning message (1) or
not (0)
1
N
Y
Critical range text
0 – 100 characters
“Out of
Critical
Range”
N
Y
Normal range HI limit
10 – 600 mg/dL
0.6 – 33.3 mmol/L
600 mg/dL
33.3 mmol/L
Y
Y
Normal range LO limit
10 – 600 mg/dL
0.6 – 33.3 mmol/L
10mg/dL
0.6 mmol/L
Y
Y
STAT tests allowed
0: no
1: yes
0
Y
Y
Number of STAT tests allowed
0 – 9
9
Y
Y
Reportable range HI limit
10 – 600 mg/dL
0.6 – 33.3 mmol/L
600 mg/dL
33.3 mmol/L
Y
Y
Reportable range LO limit
10 – 600 mg/dL
0.6 – 33.3 mmol/L
10mg/dL
0.6 mmol/L
Y
Y
Subject/Attribute
Range
Default
Device
DMS