168
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Meter reset
A meter reset should only be performed if all other reme-
dies have failed.
1
Place the meter on a level surface.
2
Press the On/Off button for at least 12 seconds and
release the button.
■
The meter powers off and on again.
■
The Roche logo is displayed.
If the Roche logo does not appear within 15 seconds,
place the meter in a base unit for a minimum of
15 minutes to recharge the battery.
■
The meter performs a system check. During the
system check, the
Versions
screen appears for a few
seconds, then disappears. The
Versions
screen
shows the version numbers of the meter compo-
nents. (The
Versions
screen shown here is for illus-
tration purposes only. Version numbers on your
meter may differ.)
■
A message appears informing you that the date and
time settings have been lost due to the meter reset.
3
Confirm the message
Loss of Date/Time
.
Before performing any further tests, you must now
either
enter date/time manually and then synchronize
the meter,
or
contact your system administrator for
help.
To enter date/time manually and synchronize the meter,
follow the instructions on the next page.
Kernel:
RootFS:
App:
Boot:
Build:
Versions
04.01.00
04.01.00
02.12.00
02.04.02
02.09.02
Checking system...
Performing self-checks...
Performing self-checks...
W-1302: D/T settings lost
due to meter reset or loss
of power. Before
performing further tests:
Enter D/T manually and
synchronize meter. OR: Ask
System Admin. for help.
Loss of Date/Time