Appendix •
A
185
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Operator ID
Operator ID entry control
0: None
1: Prompt
2: Scan only
3: Prompt (numeric only)
1
Y
Y
Operator ID entry control on Glucose
control only
0: yes (on controls only)
1: no (always)
1
Y
Y
Operator ID validation
(Allowed characters: a-z, 0-9,
“.” [period], “-” [hyphen])
0: none
1: length
2: list
3: list & password
0
Y
Y
Operator ID maximum length (used to
validate an operator ID if operator ID
validation mode is set to “Length”)
0 – 20
20
Y
Y
Operator ID minimum length (used to
validate an operator ID if operator ID
validation mode is set to “Length”)
0 – 20
0
Y
Y
Operator ID timeout (determines the time in
sec. before the operator is logged out after
power-off). 0 = immediately logged out.
0 – 3,600sec
0
N
Y
Operator password length
4-20 characters (only a-z,
0-9 allowed)
N/A
N
N
Operator certification expiry warning
0-90 days
0 (=off)
N
Y
Operator Name length
0-25 characters
N/A
N
N
Operator ID and patient ID barcode masks
Operator ID barcode mask (see separate
table at the end of this chapter)
0 – 300 characters*
“”
N
Y
Patient ID barcode mask (see separate
table at the end of this chapter)
0 – 300 characters*
“”
N
Y
* The number of characters in the unmasked barcode may not exceed 20.