Overview of the Accu-Chek Inform II Base Unit Hub
The Accu-Chek Inform II Base Unit Hub is able to con-
nect up to 4 Accu-Chek Inform II Base Units Light (RJ25)
and supports communication to a data management
system via Ethernet (RJ45).
The hub also provides power for the Accu-Chek Inform II
Base Unit Light. When connected to a hub an Accu-Chek
Inform II Base Unit Light does not require a separate
power supply. The LED colors of the hub are the same as
for the Accu-Chek Inform II Base Unit Light.
17
Status LED (lights up when power is connected)
19
Power input jack for the power supply unit
20
Network connection — Ethernet/RJ45 port
21
USB port
24
Connections for up to 4 Accu-Chek Inform II Base
Units Light (RJ25, data transfer, and power)
25
RJ25 cable — connects the hub to an Accu-Chek
Inform II Base Unit Light
For additional information about the Accu-Chek Inform II
Base Unit or Accu-Chek Inform II Base Unit Hub, refer to
the Accu-Chek Inform II Base Unit Hub Operator’s
Manual or the “Technical Note” stored on the Accu-Chek
Inform II Base Unit as a PDF file (see page 178 of this
manual). Or contact Accu-Chek Customer Care for tech-
nical assistance.
NOTICE
Cable mismatch
Do not insert the RJ25 cable used to connect an
Accu-Chek Inform II Base Unit Light to the hub into the
network connection (Ethernet/RJ45 port) of an
Accu-Chek Inform II Base Unit (new or legacy version).
Inserting the wrong cable into the Ethernet/RJ45 port
may cause severe damage to the Accu-Chek Inform II
Base Unit.