OMNISTIM
®
FX
2
PRO PROGRAM MENU
Pain Control
A.
IFC SENSORY
This program is used for symptomatic relief of superficial and deep pain from localized dermatome or
segmental origin. Sensory stimulation activates A-beta fibers causing the release of spinal Enkephalin and
Dynorphin to block pain at the segmental level (Gate Control). Duration of relief is typically from 30
minutes to 2 hours. Fast onset of relief usually within 15-20 minutes. Target tissue is superficial and deep.
Bilateral, bipolar or quadripolar application through the painful area or over the involved spinal segments.
Apply parallel to incision line for post operative pain management. The intensity should be set to elicit a
pleasant tingling sensation, just below a muscle contraction.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter
Set Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
N
CARRIER FREQ
5.0 (kHz)
N
BURST FREQUENCY
100 (Hz)
N
BURST RATE SCAN
20 (%)
N
RATE SCAN
ON
N
RATE SCAN TIME
20 (s)
Y
2
VECTOR
Slow 45, Fast 45, Slow 90,
Fast 90, OFF
Fast 90
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
B.
TENS-LVPC SENSORY
This program is used for symptomatic relief of superficial pain from localized dermatome or segmental
origin. Sensory stimulation activates A-beta fibers causing the release of spinal Enkephalin and Dynorphin
to block pain at the segmental level (Gate Control). Duration of relief is typically from 30 minutes to 2
hours. Fast onset of relief usually within 15-20 minutes. Target tissue is superficial. Bilateral bipolar
application through the painful area or over the involved spinal segments. Apply parallel to incision line
for post operative pain management. The intensity should be set to elicit a pleasant tingling sensation, just
below a muscle contraction.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter
Set Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y
2
PHASE DURATION
40…300 (μs)
80 (μs)
Y
3
PULSE RATE
0.0…250 (Hz)
100 (Hz)
N
PHASE DURATION MOD
20 (%)
N
MODULATION
OFF
N
RATE SCAN TIME
4 (s)
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
N
CHANNEL MODE
A/B SIM.