NOTE:
When using the Omnistim
®
Output Tester, increase unit output intensity to 40 mA or higher in order to
significantly illuminate the LED. Also, view the LED located at the end of the output tester directly. If viewed at
an angle, the LED may be too dim to notice that it is illuminated.
INSPECT BATTERY CONTACTS / VERIFY BATTERY SWITCH SET TO RECHARGEABLE
1.
Remove batteries if they are installed.
2.
Inspect the battery contacts for any corrosion or discoloration. If corrosion or discoloration is found, perform the
following steps to clean the metal contacts:
Remove the power cord from the stimulator unit.
Hold the metal contact plate in place with one finger.
Use a small piece of sand paper to clean the surface of the metal contact thoroughly as needed. (Figure 3).
Figure 3
3.
Verify the switch inside the battery compartment is set to the RECHARGEABLE BAT position (assuming
rechargeable batteries are being used, which is recommended).
4.
Install the rechargeable batteries into the unit.
5.
Verify the battery door opens and closes
correctly. The door should lock into place when
closed.
INSPECT POWER SUPPLY OPERATION
1.
Connect the power cord to the power supply and
to a power outlet.
2.
Connect the power supply cable to the stimulator
unit (Figure 4).
Figure 4
3.
Press the power supply switch ON (if applicable, Figure 5a).
NOTE:
Newer power supply model does not have a ON/OFF switch (Figure 5b).
Cleaning the
battery contact
(piece of sand
paper shown)
Connect power
supply cable to
stimulator