B.
LVPC SURFACE MUSCLE
Program used for strength development and treatment of muscle disuse atrophy. Activation of muscle with
electrical stimulation at a high intensity for a short time with long “ON/OFF” ramps reduces atrophy and
improves strength. The patient should participate to the extent possible by contracting during the
stimulation “ON” time. Target tissues are smaller superficial muscle groups. Bipolar set-up over agonist
and antagonist muscles for target muscle. The intensity should be set to elicit a grade 2 to 4 muscle
contraction. Treatment time should be reduced based on muscle fatigue. The affected joint movement
should be restricted by holding manually, or using weights or exercise bands.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter
Set Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y
6
PHASE DURATION
40…300 (μs)
70 (μs)
Y
7
PULSE RATE
0.0…250 (Hz)
50.0 (Hz)
N
PHASE DURATION MOD
0
N
MODULATION
OFF
N
RATE SCAN TIME
0 (s)
Y
1
TREATMENT TIME
0…99 (min)
10 (min)
Y
2
9
2
B OUTPUT DELAY
0.0…9.9 (s)
0.0 (s)
Y
4
RAMP UP
0.0…9.9 (s)
2.0 (s)
Y
5
RAMP DOWN
0.0…9.9 (s)
2.0 (s)
Y
2
T-ON
0…30 (s)
10 (s)
1
Y
3
T-OFF
0…199 (s)
50 (s)
1
Y
8
CHANNEL MODE
A/B SIMULTANEOUS, A/B ALTERNATE, B DELAYED
A/B SIM.
1
When T-OFF = 0, T-ON = CONTINUOUS.
2
B OUTPUT DELAY parameter is shown only when the Channel Mode is set to B DELAYED.
Functional Re-Ed
A.
SLOW CYCLE
Program simulates slow cycling, at 42 rpm, and can be used to retrain reciprocal movement of the L.E.
with or without resistance. Electrodes are applied to both quads while sitting with hips and knees flexed.
Use a lower extremity ergometer or pedal exerciser in combination with stimulation at mild to moderate
contraction intensity for closed chain functional training. Increase resistance as patient progresses in the
rehab program.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter
Set Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y
2
PHASE DURATION
40…100 (μs)
70 (μs)
N
CYCLE TIME
1.40 (s)
N
PULSE RATE
50 (Hz)
Y
1
TREATMENT TIME
0…99 (min)
15 (min)