E.
MFAC RECIPROCAL INHIBITION
This program provides for stimulation of the spastic muscle’s antagonists. It activates reciprocal inhibition
of the spastic muscle reducing tone. Slow ramps decrease the potential to trigger spasticity of the agonist.
A single channel (A or B) is used over the spastic muscle’s antagonist. Intensity should be set to elicit a
grade 2 to 3 muscle contraction. Intensity should not be too high to induce overflow activation of the
spastic muscle. Treatment time can be reduced based on muscle fatigue and inhibition of spasticity.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter
Set Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y
2
CARRIER FREQ
2.0, 2.5, 4.0, 5.0, 10.0 (kHz)
2.5 (kHz)
Y
3
BURST FREQUENCY
0.0…250 (Hz)
50 (Hz)
N
BURST RATE SCAN
20 (%)
N
RATE SCAN
ON
N
RATE SCAN TIME
4 (s)
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
Y
2
9
2
B OUTPUT DELAY
0.0…9.9 (s)
0.0 (s)
Y
6
RAMP UP
0.0…9.9 (s)
6.0 (s)
Y
7
RAMP DOWN
0.0…9.9 (s)
4.0 (s)
Y
4
T-ON
0…30 (s)
12 (s)
1
Y
5
T-OFF
0…199 (s)
18 (s)
1
Y
8
CHANNEL MODE
A/B SIMULTANEOUS, A/B ALTERNATE, B DELAYED
A/B SIM.
1
When T-OFF = 0, T-ON = CONTINUOUS.
2
B OUTPUT DELAY parameter is shown only when the Channel Mode is set to B DELAYED.
Muscle Re-Ed
A.
MFAC SURFACE DEEP MUSCLE
Program used for strength development and treatment of muscle disuse atrophy. Activation of muscle with
electrical stimulation at a high intensity for a short time with long “ON/OFF” ramps reduces atrophy and
improves strength. The patient should participate to the extent possible by contracting during the stimulation
“ON” time. Target tissues are superficial and deep muscle groups. Bipolar set-up over agonist and antagonist
muscles for target muscle. The intensity should be set to elicit a grade 2 to 4 muscle contraction. Treatment
time should be reduced based on muscle fatigue. The affected joint movement should be restricted by holding
manually, or using weights or exercise bands.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter
Set Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y
9
CARRIER FREQ
2.0, 2.5, 4.0, 5.0, 10.0 (kHz)
2.5 (kHz)
Y
6
BURST FREQUENCY
0.0…250 (Hz)
75 (Hz)
N
BURST RATE SCAN
20 (%)
N
RATE SCAN
ON
N
RATE SCAN TIME
4 (s)
Y
1
TREATMENT TIME
0…99 (min)
10 (min)
Y
2
8
2
B OUTPUT DELAY
0.0…9.9 (s)
0.0 (s)
Y
4
RAMP UP
0.0…9.9 (s)
2.0 (s)
Y
5
RAMP DOWN
0.0…9.9 (s)
2.0 (s)
Y
2
T-ON
0…30 (s)
10 (s)
1
Y
3
T-OFF
0…199 (s)
50 (s)
1
Y
7
CHANNEL MODE
A/B SIMULTANEOUS, A/B ALTERNATE, B DELAYED
A/B SIM.
1
When T-OFF = 0, T-ON = CONTINUOUS.
2
B OUTPUT DELAY parameter is shown only when the Channel Mode is set to B DELAYED.