TECHNICAL SPECIFICATIONS
GENERAL:
Dimensions:
12" (30 cm) D x 6.5" (16.5cm) W x 3" (7.6cm) H
Weight
(Including batteries):
3lbs. 14oz. (1.75kgs)
Operating Power:
120/240VAC; 50/60Hz; 50W, 4 x 1.5VDC “D” cell alkaline or rechargeable batteries.
Battery Life:
New alkaline batteries operate the system for 40 hours at full output and over 100 hours at normal settings.
Battery voltage is displayed and monitored. With the 4.4Ah NiCad rechargeable cells the system will
operate approximately 10 hours at normal settings.
Display System:
Super Twist LCD full character display with adjustable contrast /viewing angle.
Push Buttons:
Polyester embossed overlay for tactile feel and infection control. Speed and sensitivity are fully adjustable.
System Memory:
The system remembers all prior custom settings from treatment to treatment in active battery powered
RAM memory. (Must be reset if the batteries are changed or discharged.)
System Architecture:
CMOS integrated micro-controller with on board memory and instruction set.
STIMULATION SYSTEM:
Output:
Constant voltage up to maximum preset current limit
Output Waveform:
AC sine wave into physiologic load, IFC and MFAC modes, and biphasic square wave (LVPC) or mono-
phasic pulsed current (HVPC mode). Output Amplitude: IFC FD and NMES modes-0 to 70 mA RMS
current into a 500Ω, LVPC mode 75mA P-P, HVPC mode 250VDC peak into 500Ω loads.
Channel Isolation:
Independent transformer isolation.
Line Leakage:
<50 µA when operated with the charger system.
MFAC (MEDIUM FREQUENCY ALTERNATING CURRENT) MODE:
Waveform:
Modified sine wave into physiologic load 2 KHz, 2.5 KHz, 4 KHz , 5 KHz AND 10 KHz carrier frequency.
Burst Rate:
Adjustable from 0.1 to 250bps, 50% duty factor
Rate Scan:
From 0-50%, from 0-20 seconds
LVPC (LOW VOLTAGE PULSED CURRENT) MODE:
Waveform:
Asymmetrical biphasic square wave
Phase Duration:
10 - 300 µs
Pulse Rate:
0.1 – 250 pps
IFC (INTERFERENTIAL CURRENT) MODE:
Frequency Difference Rate:
Output channel A, fixed at 2.0 KHz, 2.5, KHz, 4.0 KHz, or 5.0 KHz. Output channel B variable from
channel A fre .0 to 250 Hz. Beat rate may be set during treatment.
Full Field Burst Rate:
.1 – 250 Hz
Rate Scan:
0-20 seconds, 0-50% modulation. Upper and lower frequencies are fully programmable and sweep time is
adjustable in set mode.
Vector:
Scans amplitude of channel A relative to channel B which varies concurrently out of phase.
PENS (PATTERNED ELECTRICAL NEUROMUSCULAR STIMULATION) MODE:
Waveform:
Asymmetric Biphasic
Phase Duration:
Variable from 40 µs to 100 µs
Pulse Rate:
Set at 50 Hz