background image

Printing Test Results from a

Martel Printer or i-STAT Printer

Without Downloader or Downloader/Recharger

1.    Turn printer on if green power light is not on.

2.    Align IR windows of handheld and printer.

3.    Display results.

4.    Press          .

5.    Do not move handheld or printer until printing is complete.

6.    If printer is not powered from a wall outlet, turn printer off.

With Downloader or Downloader/Recharger

1.    Place handheld in Downloader or Downloader/Recharger that is wired to the printer.

2.    Display results.

3.    Press          .  Do not move handheld or printer until printing is complete.

4.    If printer is not powered from a wall unit using the AC adapter, turn printer off.

Printing more than one result

1.    Turn the handheld on.

2.    Press         .

3.    Press         for Data Review.

4.    Press         for List.

5.    Scroll through the test records using the 

and 

keys.

6.    Press the numbered key for the test record(s). (Press the numbered key again to

deselect a record.)

7.    Align handheld and printer IR window or place in Downloader or

Downloader/Recharger attached to printer.  Press         .

8.    Do not move handheld or printer until printing is complete.

9.    If printer is not powered from a wall unit using the AC adapter, turn printer off.

Analyzer

n

  

Storage/Transport temperature: -10 to 46 °C (14-115 °F).

n

  

The handheld’s operating temperature range is 16 to 30 °C (61-86 °F).

n

   Store handhelds near the testing location or in an area close to the temperature of

the testing area.  Do not store handhelds near equipment that gives off heat or in
direct sunlight.

Storage Conditions and Preparation for Use

Cartridges

n

  

Store at temperatures between 2 and 8 °C (35-46 °F).  Do not use after expiration
date on cartridge pouch and box.

n

  

Equilibrate a single cartridge for 5 minutes or a box of cartridges for 1 hour at room
temperature before opening pouches.

n

  

Store cartridges at room temperature for the timeframe indicated on the cartridge box.
Mark the cartridge box or cartridge pouches with the room temperature expiration date.
Do not expose to temperatures above 30 °C (86 °F).  Do not return cartridges to the
refrigerator after room temperature equilibration.

n

  

Use cartridge immediately after opening pouch.  If the pouch has been punctured,
the cartridge should not be used.

2

7

MENU

12

9

Transmitting Results using a Downloader or

Downloader/Recharger

Note:

For analyzers containing Wireless Module FCC ID: P1405W, users must install a
LAN Cable Filter (Abbott L/N 06F23-63) on all Downloader/ Rechargers that are used
in conjunction with the i-STAT1 Wireless Analyzer.  See the “Procedure for Using the i-
STAT1 Wireless Analyzer” Technical Bulletin for full details. For analyzers containing
Wireless Module FCC ID: YOPGS1500M, the LAN Cable Filter is not required.

1.    Place handheld in Downloader or Downloader/Recharger.  A

“Communication in Progress” message will appear on the
handheld display.

2.    Do not move handheld until “Communication in Progress”

message disappears.  Once the message disappears, the
transmission is successful.

      

Note:

When wireless is enabled and configured on an 
i-STAT 1 Wireless Analyzer, wireless communication will be
attempted first. If wireless communication fails, wired
communication will then be attempted.

Summary of Contents for i-STAT

Page 1: ...AT User Guide Emergo Europe Molenstraat 15 2513 BH The Hague The Netherlands Symbols Symbols can be useful in reducing the necessity for translating important information into multiple languages parti...

Page 2: ...X X where X is one digit Analyzers whose label contains two blue triangles one in each of the upper corners are referred to as having Wireless Module FCC ID YOPGS1500M without SHA 256 support Firmware...

Page 3: ...for pH PCO2 PO2 TCO2 and ionized calcium should be tested within 10 minutes Test other analytes within 30 minutes Avoid drawing air into syringes for blood gas and ionized calcium tests If not tested...

Page 4: ...lood or plasma samples collected in syringes or evacuated tubes containing lithium or sodium heparin and lled to capacity or b non heparinized whole blood or plasma samples tested within one minute of...

Page 5: ...efore scanning the barcode 1 Press to turn on handheld 2 Press i STAT Cartridge 3 Follow handheld prompts 4 Scan the lot number on the cartridge pouch Position barcode 3 9 inches from scanner window o...

Page 6: ...rge Compartment Placing a rechargeable battery into the recharging compartment will automatically initiate trickle recharging The indicator light near the recharging compartment will be green when a r...

Page 7: ...s Represents the 0 to 99 range of results Creatinine Kinase MB ng mL g L 0 0 150 0 0 0 3 5 CK MB Represents the 0 to 95 range of results Type Natriuretic pg mL ng L 15 5000 15 50 Peptide BNP Represent...

Page 8: ...r test may be used in the veri cation This is not a manufacturer s system instruction It is a suggestion to comply with regulatory requirements that may pertain to your laboratory Test Flags and Opera...

Page 9: ...sunlight Storage Conditions and Preparation for Use Cartridges n Store at temperatures between 2 and 8 C 35 46 F Do not use after expiration date on cartridge pouch and box n Equilibrate a single cart...

Page 10: ...ote If there are unsent results remaining in the handheld at the completion of a transmission attempt a Communication Ended message will appear on the handheld indicating how many unsent results are l...

Reviews: