Reviewing Test Results

n

    

The 0 key can be used to backlight the display to view
results in dim lighting.  (The backlight turns off after 90
seconds or when the 0 key is pressed again.)

n

    

Test results are displayed numerically and with bar
graphs. Tick marks indicate the reference ranges on
the bar graphs.  (Blood gases and their associated
calculated values are not displayed with bar graphs
and reference ranges.)

n

    

Test results are displayed for 2 minutes or a
customized time.  To recall the last set of results to the
screen, turn the analyzer on and press 1 for Last
Result.

n

    

To review results from the same patient, when results
are displayed, press 1 for Test Options and then 3 for
History.  Scroll through test records using the 1 and 2
keys.

n

    

To review another patient’s results, turn the analyzer
on and press the Menu key followed by the 2 key for
Data Review and the 1 key for patient.  Scan or enter
the Patient’s ID number.  Use the 1 and 2 keys to
scroll through the test records.  Or, press the Menu
key followed by the 7 key for List.  Select the test
record(s) to be reviewed and press the Enter key.

6

15

Reportable and Reference Ranges

Measured:

Reportable

Reference Range

Test

Units

Range

(arterial)

(venous)

Sodium/Na

mmol/L (mEq/L)

100 – 180

138 – 146

138 – 146

Potassium/K

mmol/L (mEq/L)

2.0 – 9.0

3.5 – 4.9

3.5 – 4.9

Chloride/Cl

mmol/L (mEq/L)

65 – 140

98 – 109

98 – 109

Glucose/Glu

mmol/L

1.1 – 38.9

3.9 – 5.8

3.9 – 5.8

mg/dL

20 – 700

70 – 105

70 – 105

g/L

0.20 – 7.00

0.70 – 1.05

0.70 – 1.05

Lactate/Lac

mmol/L

0.30 – 20.00

0.36 – 1.25

0.90 – 1.70

mg/dL

2.7 – 180.2

3.2 – 11.3

8.1 – 15.3

Creatinine/Crea

mg/dL

0.2 – 20.0

0.6 – 1.3

0.6 – 1.3

µmol/L

18 – 1768

53 – 115

53 – 115

Charging Rechargeable Battery in External Recharge Compartment

Placing a rechargeable battery into the recharging compartment will automatically initiate
trickle recharging.  The indicator light near the recharging compartment will be green when a
rechargeable battery is placed in the compartment.

STEP / ACTION

1.    The battery pack has two labels:  one for orientation in the handheld and one for

orientation in the Downloader/Recharger. With the label with the Downloader facing
up and the electrical contact end of the pack facing the contacts in the battery
compartment, insert the pack into the compartment as shown on the label.

2.    To remove the battery after it is charged, back the pack out of the compartment.

Full recharge from a discharged state takes approximately 40 hours.

Caution

If you are using rechargeable batteries, use 

only

rechargeable batteries and recharging

equipment supplied by your i-STAT distributor.  Other batteries and rechargers may affect test
results and pose other hazards to operators and patients.

Hardware Procedures

Replacing Batteries

1.    Slide the battery compartment door off.

2.    Tilt the handheld slightly to slide out the battery

carrier.

3.    Remove the old batteries from the carrier and replace

with 2 new 9V lithium batteries. 
(Ultralife Battery Part No. U9VL-J-P)  

4.    Insert the carrier back into the compartment – label

facing up and electrical contacts first.

5.    Slide the battery compartment door into place.

Charging the Rechargeable Battery

Placing a handheld in a Downloader/Recharger will automatically initiate recharging of
the rechargeable battery.  The indicator light on top of the Downloader/Recharger will be
green (trickle charge), red (fast charge), or blinking red (fast charge pending) when a
handheld with a rechargeable battery is placed in the Downloader/Recharger.  

No damage will be caused if a handheld with disposable batteries installed is placed in
the Downloader/Recharger.

Summary of Contents for i-STAT

Page 1: ...AT User Guide Emergo Europe Molenstraat 15 2513 BH The Hague The Netherlands Symbols Symbols can be useful in reducing the necessity for translating important information into multiple languages parti...

Page 2: ...X X where X is one digit Analyzers whose label contains two blue triangles one in each of the upper corners are referred to as having Wireless Module FCC ID YOPGS1500M without SHA 256 support Firmware...

Page 3: ...for pH PCO2 PO2 TCO2 and ionized calcium should be tested within 10 minutes Test other analytes within 30 minutes Avoid drawing air into syringes for blood gas and ionized calcium tests If not tested...

Page 4: ...lood or plasma samples collected in syringes or evacuated tubes containing lithium or sodium heparin and lled to capacity or b non heparinized whole blood or plasma samples tested within one minute of...

Page 5: ...efore scanning the barcode 1 Press to turn on handheld 2 Press i STAT Cartridge 3 Follow handheld prompts 4 Scan the lot number on the cartridge pouch Position barcode 3 9 inches from scanner window o...

Page 6: ...rge Compartment Placing a rechargeable battery into the recharging compartment will automatically initiate trickle recharging The indicator light near the recharging compartment will be green when a r...

Page 7: ...s Represents the 0 to 99 range of results Creatinine Kinase MB ng mL g L 0 0 150 0 0 0 3 5 CK MB Represents the 0 to 95 range of results Type Natriuretic pg mL ng L 15 5000 15 50 Peptide BNP Represent...

Page 8: ...r test may be used in the veri cation This is not a manufacturer s system instruction It is a suggestion to comply with regulatory requirements that may pertain to your laboratory Test Flags and Opera...

Page 9: ...sunlight Storage Conditions and Preparation for Use Cartridges n Store at temperatures between 2 and 8 C 35 46 F Do not use after expiration date on cartridge pouch and box n Equilibrate a single cart...

Page 10: ...ote If there are unsent results remaining in the handheld at the completion of a transmission attempt a Communication Ended message will appear on the handheld indicating how many unsent results are l...

Reviews: