Message

Cause

Action

Sample Positioned
Short of Fill Mark

Cartridge underfilled.

Use another cartridge - fill to Fill
Mark.

Sample Positioned
Beyond Fill Mark

Cartridge overfilled.

Use another cartridge - do not fill
beyond Fill Mark.

Test Cancelled by
Operator

User did not respond
to mandatory prompt
before analyzer time
out.

No action required.

Cartridge Type Not
Recognized

Software does not
recognize cartridge.

Update software. Check to see if
cartridges are expired.

Analyzer Error, Use
Electronic
Simulator

Handheld detects
problem from which it
is likely to recover.

Insert the Electronic Simulator. If
PASS, continue to use handheld.

Analyzer Error, 
See Manual

Handheld detects
problem from which it
may not recover.

Insert Electronic Simulator. If PASS,
insert a cartridge with sample
or control. If the code does not
reappear, continue to use handheld.

No display

Disposable batteries
dead or rechargeable
battery fully
discharged. Keypad
not responding. Start
switch broken.

Change disposable batteries or
recharge battery. If still no display,
call Support Services.

Cartridge Locked
does not disappear
after test cycle
completed

Dead battery(s).
Mechanical
problem.

Wait until handheld turns off. Turn
handheld on. If resets, remove
cartridge. If not, change or recharge
battery(s) and turn handheld on.

Quality Check Messages and Codes 

(continued)

Blood Collection

Acceptable Samples for Cartridges

n

     Arterial:  Plain syringe, heparinized syringe labeled for analytes to be tested and filled to

capacity, or syringe with minimum volume of heparin to prevent clotting (10 U/mL of blood).
For ionized calcium, use balanced heparin syringes.  Mix heparinized syringes by rolling
between palms for at least 5 seconds in two directions, then invert the syringe repeatedly
for at least 5 seconds.  Test for lactate immediately.  Samples for pH, 

P

CO

2

P

O

2 , 

TCO

and

ionized calcium should be tested within 10 minutes.  Test other analytes within 30 minutes.

Avoid drawing air into syringes for blood gas and ionized calcium tests.

If not tested immediately, remix and discard 2 drops of blood before filling cartridge.

Do not use iced samples.

n

     Venous:  Collection tube with lithium heparin filled to capacity and mixed by gentle

inversion at least 10 times.  Test within 10 minutes.

Do not leave tourniquet on for more than 2 minutes.

Do not draw above an I.V.

n

     Skin puncture:  Lithium heparin capillary tubes for testing all analytes except ionized

calcium.  For all analytes, including ionized calcium, use plain or balanced heparin
capillary tubes.  Test immediately.

Allow alcohol to dry over puncture site before collecting sample.

Do not “milk” finger or heel while collecting sample.

n

     Coagulation Tests:

The ACT test may be performed using venous or arterial samples, while the PT/INR
test may be performed using capillary or venous samples.

Use plain plastic syringes or plastic evacuated tubes with no anticoagulant,
activators, or serum separators.

Test sample immediately upon draw.

For venipuncture, some experts recommend drawing and discarding a sample of
at least 1 mL prior to drawing samples for coagulation testing.

If a second measurement is needed, draw a fresh sample.

For indwelling line testing for ACT:

a.    Fluid drip through the line must be discontinued.

b.    If blood must be drawn from an indwelling line, possible heparin contamination

and specimen dilution should be considered. The line should be flushed with 
5 mL of saline and the first 5 mL of blood or six dead space volumes should be
discarded.

c.    Withdraw the sample into a fresh plastic syringe with no anticoagulant and

test immediately.

For extracorporeal line testing for ACT:

a.    Flush the extracorporeal blood access line by withdrawing 5 mL of blood

into a syringe and discard the syringe.

b.    Withdraw the sample into a fresh plastic syringe with no anticoagulant and

test immediately.

18

3

Summary of Contents for i-STAT

Page 1: ...AT User Guide Emergo Europe Molenstraat 15 2513 BH The Hague The Netherlands Symbols Symbols can be useful in reducing the necessity for translating important information into multiple languages parti...

Page 2: ...X X where X is one digit Analyzers whose label contains two blue triangles one in each of the upper corners are referred to as having Wireless Module FCC ID YOPGS1500M without SHA 256 support Firmware...

Page 3: ...for pH PCO2 PO2 TCO2 and ionized calcium should be tested within 10 minutes Test other analytes within 30 minutes Avoid drawing air into syringes for blood gas and ionized calcium tests If not tested...

Page 4: ...lood or plasma samples collected in syringes or evacuated tubes containing lithium or sodium heparin and lled to capacity or b non heparinized whole blood or plasma samples tested within one minute of...

Page 5: ...efore scanning the barcode 1 Press to turn on handheld 2 Press i STAT Cartridge 3 Follow handheld prompts 4 Scan the lot number on the cartridge pouch Position barcode 3 9 inches from scanner window o...

Page 6: ...rge Compartment Placing a rechargeable battery into the recharging compartment will automatically initiate trickle recharging The indicator light near the recharging compartment will be green when a r...

Page 7: ...s Represents the 0 to 99 range of results Creatinine Kinase MB ng mL g L 0 0 150 0 0 0 3 5 CK MB Represents the 0 to 95 range of results Type Natriuretic pg mL ng L 15 5000 15 50 Peptide BNP Represent...

Page 8: ...r test may be used in the veri cation This is not a manufacturer s system instruction It is a suggestion to comply with regulatory requirements that may pertain to your laboratory Test Flags and Opera...

Page 9: ...sunlight Storage Conditions and Preparation for Use Cartridges n Store at temperatures between 2 and 8 C 35 46 F Do not use after expiration date on cartridge pouch and box n Equilibrate a single cart...

Page 10: ...ote If there are unsent results remaining in the handheld at the completion of a transmission attempt a Communication Ended message will appear on the handheld indicating how many unsent results are l...