Message
Cause
Action
Date Invalid,
Check Clock
Date outside six
month lifetime of
software.
Select 5-Clock Set from
Administration Menu. (Password
protected.)
Dead Batteries,
Replace Batteries
Insufficient power to
complete a test cycle.
Replace disposable batteries or
recharge the rechargeable battery.
Temperature Out of
Range, Check
Status page
Temperature outside
operating range of 16
to 30 °C.
Check handheld temperature by
pressing 1 for Analyzer Status under
the Administration Menu. Move
handheld to warmer area if below
operating range or to cooler area if
above the range.
Invalid or Expired
CLEW
Software expired or
corrupt.
Verify that the handheld’s date is
correct. Change software if expired.
Update again if not expired.
Analyzer Interrupted,
Use Another
Cartridge
Last cartridge run not
completed.
Check that the battery pack is inserted
properly. Check for Low Battery
startup warning.
Cartridge Error
Usually problem with
sample or cartridge
filling.
Use another cartridge. If same code
repeats more than twice, try another
handheld.
Cartridge Preburst
Calibrant pack burst
before cartridge inserted
into handheld.
Use another cartridge - do not press
on center of cartridge. Check that
cartridges have not been frozen.
Unable to Position
Sample
Cartridge not sealed.
Clot in sample.
Aberrant cartridge.
Use another cartridge.
n
CHEM8+ Cartridges
a. CHEM 8+ cartridges require the use of whole blood collected in non-
heparinized evacuated tubes or syringes, as long as sample is tested
immediately upon draw, or;
b. Heparinized whole blood collected in balanced heparin syringes or capillary
tubes, or;
c. Heparinized whole blood collected in evacuated tubes containing lithium
heparin, as long as the tubes are filled to capacity.
n
Troponin I/cTnI and CK-MB Tests
•
cTnI and CK-MB cartridges require the use of either:
a. heparinized whole blood or plasma samples collected in syringes or
evacuated tubes containing lithium or sodium heparin and filled to capacity, or
b. non-heparinized whole blood or plasma samples tested within one minute
of drawing from a patient into a plastic syringe or plastic evacuated tube
containing no additives.
•
The use of whole blood or plasma samples containing other anticoagulants such as
EDTA, oxalate, and citrate will cause deactivation of the alkaline phosphatase,
resulting in decreased cTnI or CK-MB readings.
•
Capillary tubes and direct skin punctures (e.g., fingersticks) should not be used
with the cTnI or CK-MB cartridge.
n
BNP Tests
•
BNP cartridges require the use of EDTA whole blood or plasma samples collected
in plastic syringes or evacuated tubes containing EDTA and filled to capacity.
•
The use of whole blood or plasma samples containing other anticoagulants such
as oxalate and citrate is not recommended.
•
Capillary tubes and direct skin punctures (e.g., fingersticks) should not be used
with the BNP cartridge.
n
Total ß-hCG Tests
•
Total ß-hCG cartridges require the use of heparinized whole blood or plasma
samples collected in plastic syringes or evacuated tubes containing lithium or
sodium heparin and filled to capacity.
•
The use of whole blood or plasma samples containing other anticoagulants such
as EDTA, oxalate and citrate will cause deactivation of the alkaline phosphatase,
resulting in decreased Total ß-hCG readings.
Limitations
Interfering substances in the patient’s sample may cause an increase or decrease in a result.
Refer to the Cartridge and Test Information Sheets and Technical Bulletins for substances
and/or conditions that may interfere with cartridge tests.
4
17
Quality Check Messages and Codes
Message
Action
Electronic Simulator Test Required
Insert Electronic Simulator.
Stored Memory Low
Place analyzer in Downloader.
Stored Memory Full
Place analyzer in Downloader.
Upload Required
Place analyzer in Downloader.
Battery Low
Replace batteries or recharge battery.
CLEW Expiring, Update Required
Update software.