background image

Troubleshooting

Unexpected Results

When results do not reflect the patient’s condition, repeat the test using a fresh cartridge
and sample.  If results are still suspect, test the lot of cartridges in use with i-STAT control
solutions.  If the controls are in range, there may be an interfering substance in the
sample.  Check the Cartridge and Test Information sheets for the test in question.  Test by
another method to verify the result.  If the controls are out of range there may be a
problem with that particular cartridge lot.  Use another lot number or repeat the test using
another method, and refer to Support Services Information in the Technical Bulletin
section of the i-STAT 1 System Manual.

Startup Messages

The handheld performs self-checks when it is turned on.  If a condition that should be
corrected in the near future, but that will not affect results, is detected, a warning is
displayed.  The operator presses the 1 key to access the Test Menu.  The analyzer can be
customized to lock out the operator until the corrective action is taken.

16

5

Replacing Paper in the i-STAT Printer

1.    Open the paper compartment lid by pulling up on the release lever and remove

any remaining paper.

2.    Reel off a few centimeters of paper from the new paper roll, with the leading

edge of the paper feeding forward from the bottom of the roll.

3.    Sit the new paper roll in the compartment such that the leading edge is resting

outside the compartment on the printer casing.

4.    Close the lid until it snaps into place.

Patient Test Procedures

Cartridge Test Procedure

• DO NOT open cartridge pouch before scanning the barcode.

1.  Press          to turn on handheld.

2.  Press          i-STAT Cartridge.

3.  Follow handheld prompts.

4.  Scan the lot number on the cartridge pouch.

Position barcode 3 - 9 inches from scanner 
window on the handheld.

Press and hold           to activate the scanner.

Align the red laser light so it covers entire barcode.

The handheld will beep when it reads the 
barcode successfully.

5.  Continue normal procedures for preparing the sample, filling and sealing cartridge.

2

SCAN

Laser Radiation – Do not stare into beam.  Class 2 laser product.
Laser Diode 650 nm Maximum Output 1.0 mW.

6.  Push the sealed cartridge into the handheld port until it clicks into place.  Wait for the test 

to complete.

    

Note: For ACT, PT/INR, Hct and immunoassay testing, the handheld must remain on a level
surface with the display facing up during testing. A level surface includes running the handheld
in the Downloader/Recharger.

7.  Review results.

Cleaning the Handheld and Downloader

Please see Section 17 of the i-STAT 1 System Manual (Routine Care of the Analyzer and
Downloader) for instructions on cleaning and decontaminating the handheld and downloader.

Replacing Paper in the Martel Printer

1.    Squeeze the front and back of the paper cup to open.

2.    Remove remaining paper by pressing the Paper feed button.  Do not pull paper

through printer mechanism.

3.    Reel off a few centimeters of paper from the new paper roll and check that the end

has a clean straight edge.

4.    Slide the leading edge of the paper through the paper entry slot until you feel

resistance.  Paper feeds from underneath the roll.

5.    Press the Paper Feed button and feed paper through the printer mechanism.

6.    Keep the Paper Feed button depressed until the paper passes through the paper 

exit slot.

7.    Sit the new paper roll in the paper cup and close the lid.

Summary of Contents for i-STAT

Page 1: ...AT User Guide Emergo Europe Molenstraat 15 2513 BH The Hague The Netherlands Symbols Symbols can be useful in reducing the necessity for translating important information into multiple languages parti...

Page 2: ...X X where X is one digit Analyzers whose label contains two blue triangles one in each of the upper corners are referred to as having Wireless Module FCC ID YOPGS1500M without SHA 256 support Firmware...

Page 3: ...for pH PCO2 PO2 TCO2 and ionized calcium should be tested within 10 minutes Test other analytes within 30 minutes Avoid drawing air into syringes for blood gas and ionized calcium tests If not tested...

Page 4: ...lood or plasma samples collected in syringes or evacuated tubes containing lithium or sodium heparin and lled to capacity or b non heparinized whole blood or plasma samples tested within one minute of...

Page 5: ...efore scanning the barcode 1 Press to turn on handheld 2 Press i STAT Cartridge 3 Follow handheld prompts 4 Scan the lot number on the cartridge pouch Position barcode 3 9 inches from scanner window o...

Page 6: ...rge Compartment Placing a rechargeable battery into the recharging compartment will automatically initiate trickle recharging The indicator light near the recharging compartment will be green when a r...

Page 7: ...s Represents the 0 to 99 range of results Creatinine Kinase MB ng mL g L 0 0 150 0 0 0 3 5 CK MB Represents the 0 to 95 range of results Type Natriuretic pg mL ng L 15 5000 15 50 Peptide BNP Represent...

Page 8: ...r test may be used in the veri cation This is not a manufacturer s system instruction It is a suggestion to comply with regulatory requirements that may pertain to your laboratory Test Flags and Opera...

Page 9: ...sunlight Storage Conditions and Preparation for Use Cartridges n Store at temperatures between 2 and 8 C 35 46 F Do not use after expiration date on cartridge pouch and box n Equilibrate a single cart...

Page 10: ...ote If there are unsent results remaining in the handheld at the completion of a transmission attempt a Communication Ended message will appear on the handheld indicating how many unsent results are l...