background image

14

7

Measured 

(continued)

:

Reportable

Reference Range

Test

Units

Range

(arterial)

(venous)

pH

6.50 – 8.20

7.35 – 7.45

7.31 – 7.41

P

CO

2

mmHg

5 – 130

35 – 45

41 – 51

kPa

0.67 – 17.33

4.67 – 6.00

5.47 – 6.80

TCO

2

mmol/L (mEq/L)

5-50

23 – 27

24 – 29

(on the CHEM8+ cartridge only)

P

O

2

mmHg

5 – 800

80 – 105

kPa

0.7 – 106.6

10.7 – 14.0

Ionized Calcium/iCa

mmol/L

0.25 – 2.50

1.12 – 1.32

1.12 – 1.32

mg/dL

1.0 – 10.0

4.5 – 5.3

4.5 – 5.3

Urea Nitrogen/BUN 

mg/dL

3 – 140

8 – 26

8 – 26

Urea

mmol/L

1 – 50

2.9 – 9.4

2.9 – 9.4

mg/dL

6 – 300

17 – 56

17 – 56

Hematocrit/Hct

%PCV

15 – 75

38 – 51

38 – 51

Fraction

0.15 – 0.75

0.38 – 0.51

0.38 – 0.51

Celite

®

Activated

seconds

50 – 1000

74 – 125 (Prewarm)

74– 125 (Prewarm)

Clotting Time/

84 – 139 (Nonwarm)

84 – 139 (Nonwarm)

Celite

ACT

The range from 80 - 1000 seconds has been verified through method comparison studies.

Kaolin Activated

seconds

50 – 1000

74 – 137 (Prewarm)

74 – 137 (Prewarm)

Clotting Time/

82 – 152 (Nonwarm)

82 – 152 (Nonwarm)

Kaolin

ACT

The range from 77 - 1000 seconds has been verified through method comparison studies.

Prothrombin Time/PT

INR

0.9 – 8.0

Performance characteristics have not been established for INRs above 6.0.

Troponin I / cTnI

ng/mL (µg/L)

0.00 – 50.0

0.00 – 0.03*

0.00 – 0.08**

Performance characteristics have not been established for cTnI values above 35.00 ng/mL.
* Represents the 0 to 97.5% range of results.
** Represents the 0 to 99% range of results.

Creatinine Kinase MB/

ng/mL (µg/L)

0.0 – 150.0

0.0 – 3.5***

CK-MB

***Represents the 0 to 95% range of results.

β

-Type Natriuretic

pg/mL (ng/L)

15 – 5000

<15 – 50#

Peptide / BNP

# Represents the 0 to 95% range of results.

Total Beta-Human

IU/L

5.0 – 2000.0

<5.0

Chorionic Gonadotropin
/ß-hCG

Celite is a registered trademark of Celite Corporation, Santa Barbara, CA. for its diatomaceous earth products.

Procedure

1.    Turn the handheld on and press         to access the Administration Menu.

2.    Press         for Quality Tests.

3.    Press         for Control.

4.    Press         for i-STAT Cartridge.

5.    Scan or enter Operator ID.

6.    Enter the control lot number.

7.    Enter the cartridge lot number.

8.    Fill a cartridge with the control and close the cover.

9.    Insert the cartridge into the cartridge port.

10.  Enter chart page information if applicable.

11.  View results on handheld’s display.

12.  Remove and discard cartridge when Cartridge Locked message disappears.

13.  Press         for Test Options on the results page and press         for Next Level if

testing another level of control. 

3

1

1

1

1

MENU

Verification of Cartridges Using 2 Levels of Specified Controls

Cartridge Type                                                              Control Type

G, Crea, G3+, CG4+

TriControls 

or

i-STAT controls  

CHEM8+, E3+, EC4+, 6+

TriControls

EC8+, EG6+, EG7+, CG8+

ACTk, ACTc

i-STAT ACT controls 

PT/INR

i-STAT PT/INR controls 

cTnI

i-STAT cTnI controls 

CK-MB

i-STAT CK-MB controls 

BNP

i-STAT BNP controls

ß-hCG

i-STAT ß-hCG control

n

  

Verify that the storage conditions listed above have been maintained.

Summary of Contents for i-STAT

Page 1: ...AT User Guide Emergo Europe Molenstraat 15 2513 BH The Hague The Netherlands Symbols Symbols can be useful in reducing the necessity for translating important information into multiple languages parti...

Page 2: ...X X where X is one digit Analyzers whose label contains two blue triangles one in each of the upper corners are referred to as having Wireless Module FCC ID YOPGS1500M without SHA 256 support Firmware...

Page 3: ...for pH PCO2 PO2 TCO2 and ionized calcium should be tested within 10 minutes Test other analytes within 30 minutes Avoid drawing air into syringes for blood gas and ionized calcium tests If not tested...

Page 4: ...lood or plasma samples collected in syringes or evacuated tubes containing lithium or sodium heparin and lled to capacity or b non heparinized whole blood or plasma samples tested within one minute of...

Page 5: ...efore scanning the barcode 1 Press to turn on handheld 2 Press i STAT Cartridge 3 Follow handheld prompts 4 Scan the lot number on the cartridge pouch Position barcode 3 9 inches from scanner window o...

Page 6: ...rge Compartment Placing a rechargeable battery into the recharging compartment will automatically initiate trickle recharging The indicator light near the recharging compartment will be green when a r...

Page 7: ...s Represents the 0 to 99 range of results Creatinine Kinase MB ng mL g L 0 0 150 0 0 0 3 5 CK MB Represents the 0 to 95 range of results Type Natriuretic pg mL ng L 15 5000 15 50 Peptide BNP Represent...

Page 8: ...r test may be used in the veri cation This is not a manufacturer s system instruction It is a suggestion to comply with regulatory requirements that may pertain to your laboratory Test Flags and Opera...

Page 9: ...sunlight Storage Conditions and Preparation for Use Cartridges n Store at temperatures between 2 and 8 C 35 46 F Do not use after expiration date on cartridge pouch and box n Equilibrate a single cart...

Page 10: ...ote If there are unsent results remaining in the handheld at the completion of a transmission attempt a Communication Ended message will appear on the handheld indicating how many unsent results are l...

Reviews: