2101510MNA H | NGC82 00 AN D P GC 1000 US ER MA NUA L |
65
Figure 6-29: Run the calibration stream (Stream 4)
Figure 6-30: Calibration stream analysis running
2.
Allow the stream to process for two or three cycles (approximately 10 to 15 minutes). During
the final cycle, change the next mode to
Hold
. When the device completes the current cycle, it
will enter hold mode.
3.
Select
Next
to verify analysis results.
4.
At the
Results
tab, compare normalized percent for each component to the component and
percent listed on the calibration blend cylinder. Component percentages should be the same.