102
| NG C820 0 AN D PGC1000 US ER MAN UA L | 2101510 MNAH
To update:
1.
Click the
32 Bit loader
icon on the PCCU top menu.
Figure 10-4: Select 32-bit loader
2.
Click
Yes
at the warning if you have backed up your device data. If you have not backed up the
analyzer, see section
10.6 Back up configuration files
3.
At the Connection Setup window (
a.
Select
Network
in the Connection Type drop-down list. "Network" is the correct option,
even though the physical Ethernet is point-to-point. A point-to-point connection is
recommended for update.
b.
Type the IP address of the target analyzer in the IP Address field.
Figure 10-5: Connection Setup
4.
Click
Connect
. The loader screen displays.
IMPORTANT NOTE:
Always verify that in the Connection field, at the bottom left corner of the
loader screen, the option:
Network
displays; and that the IP address is the address of the target
analyzer. Avoid upgrading the wrong device. Direct connection to the device is always
recommended to avoid connecting to the wrong device over a network.
5.
Load the Flash (
):
a.
Click the
Flash
browse button to locate the Flash file in your system.
b.
Select the Flash file:
2102411-0xxEX.ngc
. Where xx is the version of the update package.