CoriolisMaster FCB400, FCH400
CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
99
Menu: Device Setup
Menu / parameter
Description
Device Setup
...Access Control
Selection of submenu ‘...Access Control’ using
.
...Sensor
Selection of submenu ‘...Sensor’ using
.
...Transmitter
Selection of submenu ‘...Transmitter’ using
.
...System Zero
Selection of submenu ‘...System Zero’ using
.
...Concentration
Selection of submenu ‘...Concentration’ using
.
The menu is only available if the DensiMass function is activated.
...Variable Matrix
Selection of submenu ‘...Variable Matrix’ using
.
The menu is only available if the DensiMass function is activated.
...Field Optimization
Selection of submenu ‘...Field Optimization’ using
.
Device Setup / ...Access Control
Standard Password
Entry / change of the password for the ‘Standard’ access level.
Read Only Switch
Indicator of the position of the write protection switch.
For further information, see chapter
Hardware settings
on page 64.
Device Setup / ...Sensor
Range Mode Config
Activation of the second measuring range for the mass and volume flow.
The setting can be performed separately for the mass flow rate (Qm) and volume flow (Qv). Thus you have the possibility to
quickly switch between two measuring ranges (e.g. Qm Max and Qm Max2). Switching is performed via the parameters ‘Qm
Range Mode’,’Qv Range Mode’ or via the correspondingly configured digital input.
• Disabled: Second measuring range for mass and volume flow rate deactivated.
• Qm and Qv: Second measuring range for mass and volume flow rate activated.
• Qm only: Second measuring range for mass flow activated.
• Qv only: Second measuring range for volume flow activated.
Qm Max DN
Maximum mass flow for the selected nominal diameter.
The value is set automatically via the selected nominal diameter.
Qm Max
Setting of the upper measuring range value 1 for the mass flow for forward flow and reverse flow. The value is also used to
calculate the corresponding percentage value. This parameter is only available if the mass flow output ‘Mass Flow [unit]’
was selected when configuring the power and digital outputs.
Qm Max 2
Setting of the upper measuring range value 2 for the mass flow for forward flow and reverse flow. The value is also used to
calculate the corresponding percentage value. This parameter is only available if for the parameter ‘Qm Range Mode’
the
value ‘Qm Max 2’ has been selected.
Qm Range Mode
Manual switchover between the measuring ranges (Qm Max / Qm Max 2) for the mass flow measurement. This parameter is
only available if the value Qm and Qv or Qm only has been selected for the parameter ‘Range Mode Config’.
Qv Max DN
Maximum volume flow for the selected nominal diameter.
The value is set automatically via the selected nominal diameter.
Qv Max
Setting of the upper measuring range value 1 for the volume flow for feed flow and reverse flow. The value is also used to
calculate the corresponding percentage value. This parameter is only available if the volume flow output ‘Volume Flow
[unit]’ was selected when configuring the power and digital outputs.
Qv Max 2
Setting of the upper measuring range value 2 for the volume flow for feed flow and reverse flow. The value is also used to
calculate the corresponding percentage value. This parameter is only available if for the parameter ‘Qv Range Mode’
the
value ‘Qv Max 2’ has been selected.