Menu / parameter
Description
Diagnostics / ...Output Readings
Curr.Out 31 / 32 / Uco
Display the current values and statuses of the listed inputs and outputs.
Curr.Out V1 / V2*
Curr.Out V3 / V4*
Dig.Out 41 / 42 Freq.
Dig.Out 41 / 42 State
Dig.Out 51 / 52 State
Dig.Out V1 / V2 State*
Dig.Out V3 / V4 State*
Dig.In V1 / V2 State*
Dig.In V3 / V4 State*
* Only with plug-in card present.
Diagnostics / ...Meter Erosion Mon.
Control Type
Selection of the operating mode for the erosion monitor.
—
Manual: Manual input of limit values for the erosion monitor.
—
Auto: The transmitter calculates the limit values for the erosion monitor automatically.
Factory setting: Manual.
Driver Output Max
Sets the maximum limit value for driver current.
If the driver current exceeds the limit value for the time set under the parameter ‘Driver Output Time’, alarm ‘Density too
low.Empty pipe, gas’ is triggered.
This parameter is only available if the value ‘Control Type’ has been selected for the parameter ‘Manual’.
Driver Output Time
Sets the delay time for alarm ‘Density too low.Empty pipe, gas’.
This parameter is only available if the value ‘Control Type’ has been selected for the parameter ‘Manual’.
Status Adjust
Indicates the status for automatic balancing of the erosion monitor.
—
Outstanding: The limit value is not set, erosion monitoring is not active.
—
Requested: Automatic adjustment of the erosion monitor is activated but has not yet been performed.
—
Processing: Automatic adjustment of the erosion monitor is active.
—
Done: Automatic adjustment of the erosion monitor is complete; erosion monitoring is active.
This parameter is only available if the value ‘Control Type’ has been selected for the parameter ‘Auto’.
Self Adjust Time
Sets the runtime for automatic adjustment of the erosion monitor.
The setting depends on the application and should cover several days or, if necessary, weeks.
Start Adjust
Start automatic calibration of the erosion monitor manually with
.
New Value left Time
Display the time remaining for the current automatic calibration of the erosion monitor.
Meter Erosion Level
Display the automatically calculated erosion value of the erosion monitor.
Adjusted Limit
Display the automatically calculated limit value of the erosion monitor. The limit value is calculated from the erosion value
from the automatic adjustment process and a tolerance value.
Actual Value
Displays the current erosion value for comparison with the learned limit.