CoriolisMaster FCB400, FCH400
CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
101
Menu / parameter
Description
Device Setup / ...Transmitter
...Units Selection
of
submenu ‘...Units’ using
.
...Custom Units
Selection of submenu ‘...Custom Units’ using
.
Damping Qm
Sets the damping for measuring mass flow. The value set here relates to 1
τ
(Tau). The value refers to the response time for
a stepwise mass flow change.
Damping Density
Sets the damping for measuring density. The value set here relates to 1
τ
(Tau). The value refers to the response time for a
stepwise density change.
Density Mode
Selection of operating mode for density measurement.
• Density Measured: The density of the measuring medium is calculated by the transmitter.
• Density Fixed Value: The density of the measuring medium is specified as a constant in parameter ‘Density Fixed Value’.
The ‘Density Fixed Value’ operating mode can be used to enable standard volumes to be calculated. See also chapter
Measurement of standard volumes
on page 122.
Density Fixed Value
Sets the density of the measuring medium, for example, when measuring the standard volume of gases.
...
...Cut Off
Selection of submenu ‘
…
...Cut Off’ using
.
TX Location TAG
Entry of the measuring point tag for the transmitter.
Alphanumeric, max. 20 characters
TX TAG
Entry of the TAG number for the transmitter.
Alphanumeric, max. 20 characters
Plant Data Sync.
The transmitter saves its configuration in the ‘SensorMemory’. The data is stored redundantly on the motherboard (MB) of
the transmitter and on the frontend board (FEB) of the sensor. This means the configuration can be restored quickly if any
components are replaced.
• FEB > MB: Loading the configuration from the frontend board (FEB) of the sensor.
• MB > FEB: Loads the configuration from the motherboard (MB) of the transmitter.
See also chapter
Replacing the frontend board
on page 141.
Device Reset
Restarts the device. Compensates for a short interruption of the power supply.
Restore Settings
All user-accessible parameters will be reset to the factory default settings.
...Feature Settings
Selection of submenu ‘...Feature Settings’ using
.