CoriolisMaster FCB400, FCH400
CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
123
Configuration of the standard volume measurement using the HART DTM
Example
Calculation of standard cubic meters per day (sm
3
/d)
Measured medium
Natural gas
Density in standard conditions
0.7168 kg/m
3
Measuring range
0 ... 10000 sm
3
/d
A
Set the parameter ‘Unit Volume Qv’ to ‘Custom Selectable’.
B
In parameter ‘Unit Name’, enter the desired names for the standard
volume unit ‘sm
3
/d’.
C
In parameter ‘Unit Factor’, enter the factor in l/s for the standard volume
unit ‘sm
3
/d’ (in the example 86.4 l/s).
D
Set the parameter ‘Unit Density’ to the desired density unit (in the
example kg/m
3
).
E
Set the parameter ‘Unit Density’ to the desired unit for the volume
totalizer (in the example m
3
).
F
Set the parameter ‘Unit Density’ to the desired density mode (in the
example Density fixed value).
G
In parameter ‘Density fixed value’, enter the standard density of the
measuring medium (in the example 0.71680 kg/m
3
).
G
In parameter ‘Density cut off’, enter the density for the cutoff of the
calculation (in the example 0.50 kg/m3). Adjust the parameter ‘Density
Low Check’ afterwards as needed (see page page 112).
G
Click on the ‘Apply’ button to apply the settings.
Figure 64: Screenshot from the HART DTM