116
CoriolisMaster FCB400, FCH400
CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
… 9 Operation
… Parameter descriptions
Menu / parameter
Description
Diagnostics / ...Drag Indicators / ...Sensor Indicators
Driver Output Max
Display of the maximum transmitter driver current since the last reset of the drag indicators.
Sensor Amp. Sa Min
Display of the minimum transmitter sensor amplitude since the last reset of the drag indicators.
Sensor Amp. Sb Min
Diagnostics / ...Drag Indicators / ...Temperature Indic.
Medium Min
Display of the minimum / maximum measuring medium temperature since the last reset of the drag indicators.
Medium Max
Sensor Housing Min
Display of the minimum / maximum sensor housing temperature since the last reset of the drag indicators.
Sensor Housing Max
Electr. (FEB) Min
Display of the minimum / maximum frontend board temperature since the last reset of the drag indicators.
Electr. (FEB) Max
Diagnostics / ...Simulation Mode
Simulation Switch
Manual simulation of measured values. After selecting the value to be simulated, a corresponding parameter is displayed in
the menu ‘Diagnostics / ......Simulation Mode’. The simulation value can be set here.
The output values correspond to the simulated flowrate entered.
Information ‘Configuration’ appears in the lower line of the display.
Only one measured value / output can be selected for simulation.
After power-up / restart of the device, the simulation is switched off.
Off
Curr.Out 31 / 32 / Uco
Curr.Out V1 / V2*
Curr.Out V3 / V4*
Dig.Out 41 / 42 State
Dig.Out 41 / 42 Freq.
Dig.Out 41 / 42 Pulse
Dig.Out 51 / 52 State
Dig.Out 51 / 52 Pulse
Dig.Out V1 / V2 State*
Dig.Out V3 / V4 State*
Dig.In V1 / V2 State*
Dig.In V3 / V4 State*
Qm Massflow [unit]
Qm Massflow [%]
Qv Volumeflow [unit]
Qv Volumeflow [%]
Density [unit]
Density [%]
Temperature [unit]
Temperature [%]
Hart Frequency
* Only with plug-in card present.