CoriolisMaster FCB400, FCH400
CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
111
Menu / parameter
Description
Input / Output / ...Dig.Out V1 / V2 / ...Setup Logic Output
Input / Output / ...Dig.Out V3 / V4 / ...Setup Logic Output
Logic Output Action
Selection of binary output function.
See description ‘„Input / Output / ...Dig.Out 41 / 42 / ...Setup Logic Output’.
Active Mode
Select switching properties for the binary output.
Input / Output / ...Dig.Out V1 / V2 / ...Alarm Config
Input / Output / ...Dig.Out V3 / V4 / ...Alarm Config
General Alarm
Select error messages signaled via the binary output V1 / V2 or V3 / V4.
Only if the parameter ‘Logic Output Action’ is set to Alarm Signal.
Qm Massflow Max
Qm Massflow Min
Density Min
Density Low Check
Sensor Signal Min
Driver Output Max
Input / Output / ...Dig.In V1 / V2
Input / Output / ...Dig.In V3 / V4
Function
Select a function for the digital input.
• Off:
No
function.
• Reset all Totalizer: Counter reset for all counters (forward flow, reverse flow and difference totalizer)
• Stop
all
Totalizer
:
External counter stop for all counters (forward flow, reverse flow and difference totalizer)
• Auto. Zero Adjust:Start external zero point balancing.
• Set Flowrate to zero: Sets flow measurement to 0.
• Batchflow On / Off: Start / stop fill operation (only when FillMass function is activated).
• Dual Range Mass: Switchover Qm Max / Qm Max 2.
• Dual Range Volume: Switchover Qv Max / Qv Max 2.
• Submatrix 1 or 2: Switchover for concentration matrix (only when DensiMass function is activated).
Active Mode
Select switching properties for the digital input.
Delay Time
Selection of delay time for suppressing EMC faults on the digital input.
NOTICE
If the digital input has been configured with the function ‘Batchflow On / Off’ the pulse for starting the filling process
must fit at least for the set delay time!