GAA610-M
ADVANCED EMISSION GAS MONITORING SYSTEM | OI/GAA610-M-EN REV. B
65
No.
Text
S A W F Reaction/Remark
User-Configured Messages
800
An external error occurred during:
x x
Default texts for the Message Generator function block are supplemented with the
full text defined during function block configuration.
801
A user-defined error occurred during:
x x
802
A user-defined maintenance request occurred
during
x
803
A user-defined maintenance mode event
occurred during:
x
Miscellaneous Messages
1000
This function block has an error:
x x
Is supplemented with a reference to the function block type.
1001
Condensate penetration.
Overall message for guiding reaction to condensate penetration; not logged
1002
Flow rate too high at this point!
x x
Currently not used
1003
Flow rate too low at this point!
x x
Currently not used
System Cooler
1100
Cooler temperature too high.
x x
Sample gas feed module pump is automatically turned off. Check the system cooler
and sample gas preparation system.
1101
Cooler temperature too low.
x x
1102
Condensate penetration in cooler.
x x
1103
Flow rate too low in cooler.
x
Check the system cooler and sample gas preparation system.
1104
Cooler condensate level too high.
x
1105
Cooler condensate level too high.
x
Empty the condensate bottle.
1106
Cooler reagent level too low.
x
Fill the reagent container.