![ABB CEMcaptain GAA610-M Operating Instruction Download Page 39](http://html1.mh-extra.com/html/abb/cemcaptain-gaa610-m/cemcaptain-gaa610-m_operating-instruction_3626267039.webp)
GAA610-M
ADVANCED EMISSION GAS MONITORING SYSTEM | OI/GAA610-M-EN REV. B
39
Terminal assignment – Sample gas probe
1
Heater sample gas probe
2
Status contact
Figure 27: Electrical connection spample gas probe
Electrical data
Power supply
115 to 230 V AC
Power consumption
approx. 300 VA
Heater element
PTC self limited
IP-Protection
IP 65
Operating temperature
approx. 180°C
Low temperature threshold
150 °C
Electrical connection
• Observe the cable specifications.
• Observe the ‘Interface plan’ in the drawings set.
• When routing the electrical lines, follow all applicable national
safety regulations for the installation and operation of
electrical devices.
• Follow all regulations and standards for electrical
installations on board of ships and observe the local
conditions.
• Connect the input wiring of the heated sample components
(temperature-resistant as needed).
Digital communication
Modbus® communication
Transmission of measured values and status signals as well
as analog input, digital input and digital output signals to
host systems, e.g. standard Windows applications via M-DDE
server.
Modbus slave protocol in the RTU (Remote Terminal Unit)
mode via the RS485 interface.
Figure 28: Modbus module
RS232 Interface
Version: 9-pin sub-D male connector
Pin
Signal
2
RxD
3
TxD
5
GND
RS485 interface
Version: 9-pin sub-D female connector
Pin
Signal
2
RTxD-
3
RTxD+
5
GND
Ethernet communication
Transmission of measured values and status signals as well as
analog input, digital input and digital output signals to host
systems, via Modbus/TCP.
Version: RJ45 connector, EIA/TIA 568B standard
Pin
Signal
Colour
1
Rx+
Orange/White
2
Rx
−
Orange
3
TX+
Green/White
6
TX
−
Green