![background image](http://html1.mh-extra.com/html/aandd/omniace-ra2000-series/omniace-ra2000-series_instruction-manual_3072411058.webp)
3. Setting Command – s**
Communication for RA2000 series (1WMPD4003507)
3-24
SIF Setup for pause of Input monitor
SIS Setting sync trigger
SRI Setting data acquisition and printing
Function
Sets pause of input monitor
SIF P1 (Delimiter)
Input Format
P1: Pause (0 = Cancellation, 1 = Pause)
Output Format
None
P1: When the setting of pause is omitted, operation proceeds under the
manner in which the pause has been cancelled. If the Pen Recorder mode
is set, the setting of pause is rejected.
Description
While the replay monitor screen is displayed, the monitor changes to the
input monitor screen when this command is entered.
Function
Sets sync trigger
SIS P1 (Delimiter)
Input Format
P1:Sync Trigger (0 = OFF, 1 = ON)
Output Format
None
P1: If the setting of trigger synchronization is omitted, the current setting is
held. While the Pen Recorder mode or X-Y Recorder mode is set, the
setting of Trigger synchronization = 0 is rejected.
Description
While the replay monitor screen is displayed, the monitor changes to
the input monitor screen when this command in entered.
Function
Sets data acquisition and printing
SRI P1, P2, P3, P4, P5, P6 (Delimiter)
Input Format
P1:Waveform chart Printing (0 = OFF, 1 = ON)
(Can be omitted)
P2:Memory Recording (0 = OFF, 1 = ON)
(Can be omitted)
P3:Memory recording -auto copy (0 = OFF, 1 = ON)
(Can be omitted)
P4:Memory recording -binary save (0 = OFF, 1 = ON)
(Can be omitted)
P5:HD Recording (0 = OFF, 1 = ON)
(Can be omitted)
Output Format None
Description
P1-P5 can be arbitrarily omitted. In that case, a present setting is maintained.
When all parameters are omitted, a parameter error occurs.
The setting specified by the recording mode might be disregarded.
The condition of on or off for data acquisition for each recorder mode.
Recording
Mode
Waveform
chart
printing
Memory
Recording
Memory
recording
auto copy
Memory
recording
binary
save
HD
Recording
Pen
ON
OFF
OFF
OFF
OFF
Memory
OFF
ON
*
*
OFF
HD
*
OFF
OFF
OFF
ON
Multi
*
ON
OFF
ON
ON
X-Y
*
OFF
OFF
OFF
*
ON = ON fixed, OFF = OFF fixed, * = ON/OFF through setting
The items that can be set differ depending on the recorder mode. The settings for the invalid
items are rejected.
Example: In the HD Recorder mode, on or off for chart printing is possible,
but HD acquisition cannot be set up because it is fixed to on.
Settings for the HD Recorder are rejected.
CAUTION
Summary of Contents for Omniace RA2000 series
Page 1: ...Communication Command RA2000 series INSTRUCTION MANUAL 1WMPD4003507 ...
Page 2: ......
Page 3: ...Communication Command RA2000 series INSTRUCTION MANUAL ...
Page 4: ......
Page 14: ......
Page 15: ...1 Selection of Communication Interface ...
Page 27: ...2 Overview of Communication Control ...
Page 34: ......
Page 35: ...3 Setting Command s ...
Page 63: ...4 Information Readout Command I ...
Page 90: ......
Page 91: ...5 Execution Command E ...
Page 99: ...6 File Data Operation Command F ...
Page 102: ......
Page 103: ...7 Text Operation Command T ...
Page 109: ...8 Reference ...
Page 113: ......