3. Setting Command – s**
Communication for RA2000 series (1WMPD4003507)
3-12
SCH (Set CHannel) Setting ACST amp
SAR (Set Ac strain amp R-fine)
Setting R-balance
Function
Sets ACST amp.
SCH P1, P2, P3, P4, P5, P6, P7, P8, P9 (Delimiter)
Input Format
P1: Selecting channel
P2: Amp type
P3: Input
P4: Setting range
P5: Filter
P6: Position
P7: Gage rate
P8: CAL polarity
P9: CAL level
[1-16, A] for RA2300MK
Ⅱ
and [1-32, A] for
RA2800A A means a batch setting.
4 fixed
0=OFF, 1=ON, 2=GND
2=20k
με
, 3=10k
με
, 4=5k
με
, 5=2k
με
, 6=1k
με
0=OFF, 1=10Hz, 2=30Hz, 3=100Hz, 4=300Hz
[-100.00 to 200.00] Step 0.05
[1.50 to 2.50] Step 0.01 Select 2.00 for out of
range.
0=OFF, 1=[+], 2=[-]
2=5000
με
, 3=3000
με
, 4=2000
με
, 5=1000
με
,
6=500
με
Output Format
None
Description
When P1 = A, the other selections apply to all the channels corresponding
to the selected type with P2.
When the amp type of selected channel does not correspond to P2, a
parameter error occurs.
While any action other than a chart recording is executing, an execution
error occurs.
Function
Sets R-fine (fine adjustment of resistance balance) of ACST amp
SAR P1,P2
(Delimiter)
Input Format
P1: Selecting channel
P2: Adjustment value
[1-16] for RA2300MK
Ⅱ
and [1-32] for RA2800A
[-100 to 100] can not be specified
Output Format
None
Description
After execution of the EAS command (auto balance execution), this
command adjusts the unbalanced portion.
The auto balance of ACST amp and DCST amp can be executed with the
EAS command and the EAB command respectively.
While any action other than a chart recording is executing, an execution
error occurs.
Summary of Contents for Omniace RA2000 series
Page 1: ...Communication Command RA2000 series INSTRUCTION MANUAL 1WMPD4003507 ...
Page 2: ......
Page 3: ...Communication Command RA2000 series INSTRUCTION MANUAL ...
Page 4: ......
Page 14: ......
Page 15: ...1 Selection of Communication Interface ...
Page 27: ...2 Overview of Communication Control ...
Page 34: ......
Page 35: ...3 Setting Command s ...
Page 63: ...4 Information Readout Command I ...
Page 90: ......
Page 91: ...5 Execution Command E ...
Page 99: ...6 File Data Operation Command F ...
Page 102: ......
Page 103: ...7 Text Operation Command T ...
Page 109: ...8 Reference ...
Page 113: ......