3. Setting Command – s**
Communication for RA2000 series (1WMPD4003507)
3-15
SCH (Set CHannel) Setting RMS amp
Function
Sets RMS amp.
SCH P1, P2, P3, P4, P5, P6, P7, to P15 (Delimiter)
P1: Selecting channel
P2: Amp type
P3: Input
P4: Setting range
[1-16, A] for RA2300MK
Ⅱ
and [1-32, A] for
RA2800A A means a batch setting.
9 fixed
0=OFF, 1=ON, 2=GND
The content varies depending on the P10
measurement mode.
P8=1 RMS input mode
1=350Vrms, 2=200Vrms, 3=100Vrms,
4=50Vrms, 5=20Vrms, 6=10Vrms, 7=5Vrms,
8=2Vrms, 9=1Vrms, 10=500mVrms,
11=200mVrms, 12=100mVrms
P8=2 DC input mode
1=500V, 2=200V, 3=100V, 4=50V, 5=20V,
6=10V, 7=5V, 8=2V, 9=1V, 10=500mV,
11=200mV, 12=100mV
P5: Low pass filter
P6: High pass filter
P7: Position
P8: Input mode
P9: Input combination
P10: Measurement mode
P11: Setting sensor
P12: Vibration unit
P13: Hybrid-type sensor
sensitivity
P14: Charge converter
sensitivity
P15: Acceleration sensor
sensitivity
0=OFF, 1=30Hz, 2=100Hz, 3=300Hz, 4=1kHz
0=OFF, 1=10Hz, 2=30Hz, 3=100Hz
[-100.00 to 200.00] Step 0.05
1=RMS, 2=DC
1=AC, 2=DC
0=Voltage, 1=Vibration
1=Hybrid type, 2=Standalone type
1=[m/s^2],2=[G]
[0.001 to 120.000]mV/m/s^2 or
[0.010 to 1200.00]mV/G
[0.01 to 10.0]mV/pC
[0.001 to 120.000]pC/m/s^2 or
[0.010 to 1200.00]pC/G
Input Format
The sensitivity ranges of P11 and P13 vary depending on vibration a unit.
Output Format None
Description
When P1 = A, the other selections apply to all the channels corresponding to
the selected type with P2.
When the amp type of selected channel does not correspond to P2, a
parameter error occurs.
While any action other than a chart recording is executing, an execution error
occurs.
Summary of Contents for Omniace RA2000 series
Page 1: ...Communication Command RA2000 series INSTRUCTION MANUAL 1WMPD4003507 ...
Page 2: ......
Page 3: ...Communication Command RA2000 series INSTRUCTION MANUAL ...
Page 4: ......
Page 14: ......
Page 15: ...1 Selection of Communication Interface ...
Page 27: ...2 Overview of Communication Control ...
Page 34: ......
Page 35: ...3 Setting Command s ...
Page 63: ...4 Information Readout Command I ...
Page 90: ......
Page 91: ...5 Execution Command E ...
Page 99: ...6 File Data Operation Command F ...
Page 102: ......
Page 103: ...7 Text Operation Command T ...
Page 109: ...8 Reference ...
Page 113: ......