![background image](http://html1.mh-extra.com/html/aandd/omniace-ra2000-series/omniace-ra2000-series_instruction-manual_3072411030.webp)
2. Overview of Communication Control
Communication for RA2000 series (1WMPD4003507)
2-4
2.3.
1-Byte Control Command
►
Execution is possible by sending a 1-byte control code alone, but
functions are limited. The string commands and escape sequence
command, mentioned in the preceding section, have functions of
equal or higher quality. Note that usable commands are restricted
depending on the communication interface.
●
Example of Basic Program Format
PRINT#MAD,CHR$(&H05); (MAD = Line number)
[ENQ] Outputting the status of RA2000
[CAN] Command cancel
[DC4]
Initializing RA2000
Function
Outputs the status of the RA2000.
Input Format
[ENQ](05h)
Output Format [NAK](15h): The RA2000 is operating.
[ACK](06h): The RA2000 is stopped and is waiting for a command.
Description
When the RA2000 is operating, [NAK](15h) is returned.
When the RA2000 is stopped and waiting for a command, [ACK](06) is
returned.
To see the status of the RA2000 in detail, use the
[ESC]+C command
.
Function
Cancels the command that is operating now.
Input Format
[CAN](18h)
Output Format None
Description
Command that has the same meaning as the ESP command that stops
recording.
When receiving a command, the command is canceled.
When the RA1000 is performing an operation, the operation is terminated.
However, an execution operation for amp settings such as auto-scale cannot
be terminated.
Function
Initializes the RA2000 setting.
Input Format
[DC4](14h)
Output Format None
Description
This command has the same meaning as the ESI command that initializes the
RA2000.
Execution is possible while the RA2000 is stopped.
Execution error occurs while the RA2000 is operating.
Initiali application restart is performed.
Communication settings are not initialized.
Summary of Contents for Omniace RA2000 series
Page 1: ...Communication Command RA2000 series INSTRUCTION MANUAL 1WMPD4003507 ...
Page 2: ......
Page 3: ...Communication Command RA2000 series INSTRUCTION MANUAL ...
Page 4: ......
Page 14: ......
Page 15: ...1 Selection of Communication Interface ...
Page 27: ...2 Overview of Communication Control ...
Page 34: ......
Page 35: ...3 Setting Command s ...
Page 63: ...4 Information Readout Command I ...
Page 90: ......
Page 91: ...5 Execution Command E ...
Page 99: ...6 File Data Operation Command F ...
Page 102: ......
Page 103: ...7 Text Operation Command T ...
Page 109: ...8 Reference ...
Page 113: ......