GX-A/GF-A Maintenance manual
Page 8
Corner load error
Place the specified mass at the center of the pan (1) and record the displayed value. Then
place the mass, at positions 2, 3, 4, 5 then 6. Verify that the difference between the values at
the center and at each position (the cross marks 2, 3, 4, and 5 are half the distance from the
center of the pan to the corner edge) is within the specifications.
GX-A/GF-A series
Model
Masses
Specifications
123
(Only GF-A)
100g
±0.006g
203
303
200g
±0.008g
403
±0.007g
603
300g
±0.007g
1003
500g
±0.007g
1603
±0.008g
1202
(Only GF-A)
1kg
±0.06g
2002
3002
2kg
±0.08g
4002
±0.07g
6002
3kg
±0.07g
10002
4kg
±0.10g
6001
3kg
±0.3g
10001
4kg
Weighing pan