GX-A/GF-A Maintenance manual
Page 10
Linearity / Hysteresis
(1)
GX-A/GF-A series
Linearity/ Hysteresis
After calibration by using an external weight close to the weighing capacity, place the specified
mass one by one on the pan, and check the difference between true value and display value
(linearity). After reaching weighing capacity, remove the mass one by one, and check the
difference between an increase and decrease (
hysteresis)
.
GX-A/GF-A series
Model
Masses
Linearity
Hysteresis
123
(Only GF-A)
50g×2
±0.002g
±0.002g
203
100g×2
303
100g×3
403
100g×4
603
200g×3
1003
200g×5
±0.003g
±0.003g
1603
400g×4
±0.004g
1202
(Only GF-A)
500g×2
±0.02g
±0.02g
2002
1kg×2
3002
1kg×3
4002
1kg×4
6002
2kg×3
±0.03g
±0.0
4
g
10002
2kg×5
±0.0
6
g
6001
2kg×3
±0.1g
±0.1g
10001
2kg×5