GX-A/GF-A Maintenance manual
Page 27
6.2 General Precautions
The data structure is shown below. Functions listed nearer to the bottom are more basic. If the
specific data is adjusted, all data listed above the adjusted data must also be adjusted.
Calibration
Linearity
Temperature adjustment
Corner load adjustment
Repeatability
(Reassembly of the force motor, replacement of the analog board assembly)
Analog operation range check (zero, span...D0 display)
Summary of Contents for GX-A Series