MC 

Piece Counting

 

Scale 

 
 
 

 

User Manual 

 

 
 
 

 
 
 

 
 

 

 
 

 

Summary of Contents for MC

Page 1: ...MC Piece Counting Scale User Manual ...

Page 2: ......

Page 3: ......

Page 4: ... weight 10 Parts counting 10 Automatic part weight updates 10 Setting high low limits 11 Checking Alarm 11 Accumulation 11 Automatic Accumulated Totals 11 Auto Off 1 1 Display back light 11 6 Parameter 12 Parameter block 12 Setting parameters 12 7 RS232 Protocol 18 Specification 18 Connector 18 Commands 18 Communication 18 8 Error codes 2 1 9 Drawing 22 Parts list 23 10 Dimension 24 11 Specificati...

Page 5: ... near air conditioning vents Avoid unsuitable tables The tables or floor must be rigid and not vibrate Do not place near vibrating machinery Avoid unstable power sources Do not use near large users of electricity such as welding equipment or large motors Avoid high humidity that might cause condensation Avoid direct contact with water Do not spray or immerse the scales in water Avoid air movement ...

Page 6: ...rts Description Pan Display Key board Water mark bubble On Off Switch Adaptor Jack Adjusting feet Installation Place the scale on a table Carefully place the pan to the four holes on the top cover Adjust the water mark bubble level Connect the adaptor pin to the adaptor jack Adaptor jack is locating below the right side of the scale Adaptor connects into your AC power socket Pluggable equipment mu...

Page 7: ...feet until reach center Check the level when you change the location Connect Adaptor To charge the battery insert the adaptor pin to jack Adaptor simply plug into the mains power The scale no needs to be turned on The battery should be charged for 12 hours for full capacity Right side of the display there is an LED to indicate the status of battery charging When the scale is plugged into the mains...

Page 8: ...MC Piece Counting Scale User Manual 6 3 NAMES AND FUNCTIONS Overall View Key Board ...

Page 9: ...e from the total weight and displays the result as net weight Add the current count data aggregated Also evokes the memory if pressed to balance empty Can add up to 99 values or until it reaches the maximum displayable digits Recall the accumulator memory Enter the numbers of items used for the unit weight Manually enter the weight of sample also changes the unit if they are enabled To set the upp...

Page 10: ... s illuminated the scale is in Zero point Sample Sample indicator If the scale that considers the number of samples is insufficient for an accurate count U Wt Unit weight Indicator If the Unit Weight is not sufficient for an Accurate count it will illuminate LED Battery charging status when the scale is plugged into the mains power the internal battery will be recharged If the LED is green the bat...

Page 11: ...actly zero in spite of the pan not taking any strain However you can set the display of your balance to zero any time by pressing zero key and therefore ensure that the weighing starts at zero 3 Tare The weight of any container can be tared by pressing tare button so that with subsequent weighing the net weight of the object being weighed is always displayed Load weight on the pan Press tare key Z...

Page 12: ...ue entered The sample is then stored in the balance and the net weight is displayed together with the calculated amount based on the weight per unit 3 Parts Counting Weigh a sample to calculate the unit weight to determine the average weight of items to count put a known number of pieces on the scale and insert their content The balance of the total weight divided by the number of samples and disp...

Page 13: ...will show the total weight the display Count will show the total counts aggregated and the display Unit Weight shows the number of times Articles are saved in memory These values will appear for 2 seconds The balance must return to zero or a negative value before another sample can be stored Other products may be added by pressing the M key It may take up to 99 entries or up to the capacity of the...

Page 14: ...PTION DEFAULT FI CAL Calibration F2 dI Set Division d 3000 F3 Cnt A D Count F4 AU Auto Accumulation AU oFF F5 AZn Auto Zero Range 2 d F6 Pin Set New Pin No 0000 F7 SPd Sensing Speed SPd 7 5 F8 OFF Auto Power oF 0 F9 Grv Local Gravity 9 79400 F10 bEP Beep option Low FI1 rSt Reset ...

Page 15: ...er the function by pressing the key At any time you can return to weighing b pressing the F1 CALIBRATION 1 Linear Calibration Enter the parameter menu using by password 9999 FI CAL UnLoAd Load 1 Place the first weight Load 2 Press Add the second weight Load 3 Press Add the third weight Pass Remove all weights Remove the weights The Linear Calibration procedure is finished When the calibration is f...

Page 16: ...er menu using by password Default password 0000 When the parameter menu show F1 CAL FI CAL Set Weight Load W t Remove W t F2 SET DIVISION Set the scale increment F2 di d 3000 d 6000 d 15000 Press to select d 30000 F3 A D COUNT Showing internal counts F3 Cnt XXXXX Return to menu Pass UnLoad ...

Page 17: ...4 AU ACCUMULATION Set the Accumulation Accumulate Manually Automatic Accumulation Manually send data to RS 232 Press to select F5 AZn AUTO ZERO RANGE Set the Auto Zero Range F5 AZn 0 5d 1d 2d 4d Press to select F4 AU AU oFF AU on P ASk P Cont ...

Page 18: ...in PASSWORD Set the new password pin number F6 Pin Pin 1 Enter the new four digit pin number Press F7 SPd SENSING SPEED Set the weighing speed F7 SPd SPd 7 5 SPd 15 SPd 30 Press to select SPd 60 Pin 2 Re enter the new four digit pin number Press ...

Page 19: ... is stable and not in use Auto off within three minutes Auto off within five minutes Auto off within 15 minutes Auto off within 30 minutes Press to select F9 Grv GRAVITY Set the local gravity F9 Grv 9 79400 Enter the local gravity by using numeric keys Press to select F8 oFF No Off oF 3 oF 5 oF 15 oF 30 ...

Page 20: ...EP ok Lov nG HiGH Press to select F11 rSt RESET Reset the company default settings F11 rSt SUrE Press to conform When the value is in the limit When the value is lower than the low limit When the value is in the out of the limit range When the value is higher than the limit range ...

Page 21: ......

Page 22: ...g the start command If the start command is not repeated within 30 Seconds the continuous transmission will stop until the next start command comes Communication T NO N CR LF Here the scale should always transmit the number of accumulations stored in the memory T WGT N kg CR LF Here the scale should always transmit the accumulated weight in the memory T PCS N pcs CR LF Here the scale should always...

Page 23: ...g CR LF U W 86 9018 g CR LF PCS 0pcs CR LF T NO 0 CR LF T WGT 0 0000 kg CR LF T PCS 0pcs CR LF CR LF CR LF 2 RX ST GS 1 7380 kg CR LF TARE 0 0000 kg CR LF U W 86 9018 g CR LF PCS 20pcs CR LF T NO 0 CR LF T WGT 0 0000 kg CR LF T PCS 0pcs CR LF CR LF CR LF TX M CR LF 3 RX ST GS 1 7380 kg CR LF TARE 0 0000 kg CR LF U W 86 9018 g CR LF PCS 20pcs CR LF T NO 1 CR LF T WGT 1 7380 kg CR LF T PCS 20pcs CR ...

Page 24: ...TARE 0 0000 kg CR LF U W 86 9018 g CR LF PCS 40pcs CR LF T NO 2 CR LF T WGT 5 2140 kg CR LF T PCS 60pcs CR LF CR LF CR LF 7 RX ST GS 0 0000 kg CR LF TARE 0 0000 kg CR LF U W 86 9018 g CR LF PCS 0pcs CR LF T NO 2 CR LF T WGT 5 2140 kg CR LF T PCS 60pcs CR LF CR LF CR LF TX MC CR LF 8 RX ST GS 0 0000 kg CR LF TARE 0 0000 kg CR LF U W 86 9018 g CR LF PCS 0pcs CR LF T NO 0 CR LF T WGT 0 0000 kg CR LF ...

Page 25: ...Err 5 Key board error Check the keys and connecter Err 6 A D value out of range Make sure platform empty and check the pan is installed proper Check the load cell connectors Err 9 Unstable Reading Check any air variation vibration RF noise and touching somewhere Check the load cell and connecters Err 17 Tare out of range Remove the load and restart scale again ol Over range Remove the load Re cali...

Page 26: ...MC Piece Counting Scale User Manual 22 9 DRAWING ...

Page 27: ...nector 1 Optional 14 Hexagon nut for D connector 2 Optional 15 Self thread Screw 1 20 M 3x8 16 AC adaptor jack PCB 1 17 Front display overlay 1 PC 18 Key board 1 19 Level bubble 1 14 7mm 20 Top Cover 1 ABS 21 Internal hexagon screw 4 M6x16 8 8 22 Spring Washer M6 4 HRC42 50 23 Washer M6 4 656Mn 200 300HV 24 Hexagon nut 4 M4 25 Self thread screw 4 20Mn 4x20 26 Main PCBA 1 27 Self thread screw 4 20M...

Page 28: ...kg 30kg Minimum 2g 4g 10g 20g Repeatability 0 1g 0 2g 0 5g 1g Resolution Up to 30000 Weighing Units Kg Housing ABC Plastic and Stainless Steel platform Interface RS232 Output optional Operation Temperature 0 C 40 C Power AC adaptor 9V 500mA Battery 6V 4AH Display 3x6 digits 20 mm LCD display with white LED back light Dimensions WxDxH 320 x 340 x 125mm Weight 3 8kg Internal Counts 600000 ...

Page 29: ...15 VDC 20 VDC IP Level IP65 IP64 Material Aluminum Alloy Aluminum Alloy Cable Φ 4x 35 cm Φ 8 2 four core shield Input Resistance 390Ω 15Ω 420Ω 30Ω Out put Resistance 350Ω 5Ω 350 Ω 5 Ω Temperature Range 20 C 60 C 10 C 50 C Safe overload 150 RC 150 R C Ultimate overload 200 RC 200 R C Repeatability 0 03 R O 0 02 R O Creep 0 03 R O 3 min 0 02 R O 20min Zero Balance 0 02 mV V 0 1 mV V ...