59
15. COMMANDS
15-1 Command List
Note
A command has a terminator added, that is specified using the "Terminator (
Crlf
)"
parameter of “Serial interface (
5if
)” in the function table, and is sent to the balance.
Commands to query weight data
C
Cancels
the
S
or
SIR
command.
Q
Requests the weight data immediately.
S
Requests the weight data when stabilized.
SI
Requests the weight data immediately.
SIR
Requests the weight data continuously.
E
SC
P
Requests the weight data when stabilized.
Note: The “Q” and “SI” commands, the “S” and “
E
SC
P” commands behave the same.
Commands to control the balance
CAL
Same as the CAL key.
EXC *
Calibration using an external weight.
OFF
Turns the display off.
ON
Turns the display on.
P
Same as the ON:OFF key.
PRT
Same as the PRINT key.
R
Same as the RE-ZERO key.
SMP
Same as the SAMPLE key.
T
Tare key.
Z
Same as the RE-ZERO key.
E
SC
T
Same as the RE-ZERO key.
U
Same as the MODE key.
?ID
Requests the ID number.
?SN
Requests the serial number.
?TN
Requests the model name.
?PT
Requests the tare weight.
PT:
∗∗∗∗∗
.
∗
g
Changes the tare weight.
The unit added is the current weighing unit in A&D standard format.
Note: The “R”, “Z” and “
E
SC
T” commands behave the same.
E
SC
: ASCII code 1Bh
*
:
Only for the FZ-CT series