28
9-4 Description of the Class “Environment, Display”
Condition (
Cond
)
Cond 0
This parameter is for sensitive response to the fluctuation of a weight value.
Used for powder target weighing, weighing a very light sample or when
quick response weighing is required.
After setting, the balance displays FAST.
Cond 2
This parameter is for stable weighing with slow response. Used to prevent a
weight value from drifting due to vibration or drafts.
After setting, the balance displays SLOW.
Note
With “Hold function (
Hold
)” set to “ON (
1
)”, this item is used to set the
averaging time.
Stability band width (
5t-b
)
This item controls the width to regard a weight value as a stable value. When the fluctuation per
second is less than the parameter, the balance displays the stabilization indicator and outputs the
data. The parameter influences the “Auto print mode”
5t-b 0
This parameter is for sensitive response of the stabilization indicator. Used
for exact weighing.
5t-b 2
This parameter ignores slight fluctuation of a weight value. Used to prevent
a weight value from drifting due to vibration or drafts.
Note
With “Hold function (
Hold
)” set to “ON (
1
)”, this item is used to set the
stabilization range.
Hold function (
Hold
) (Animal weighing mode)
This function is used to weigh an unstable sample such as an animal.
When the weight data is over the weighing range from zero and the display fluctuation is within the
stabilization range for a fixed period of averaging time, the processing indicator illuminates and the
balance displays the average weight of the sample. When the sample is removed from the weighing
pan, the display returns to zero automatically.
This function is available only when the hold function parameter is set to “
1
”
(the hold mode indicator
HOLD illuminates) and any weighing unit other than the counting mode is selected.
The stabilization range and averaging time are set in “Condition (
Cond
)” and “Stability band width (
5t-b
)”.
Weighing range
Averaging time
Stabilization range (Ratio of fluctuation to the weight value)
0.200 g or over
Cond 0
5t-b 0
Cond 1
5t-b 1
Cond 2
2 seconds Faster
4 seconds |
8 seconds
5t-b 2
Lesser (6.25%)
| (12.5%)
Greater (16.7%)
More
accurate
Summary of Contents for FZ-1200CT
Page 35: ...33 9 7 Data Format Examples LF...
Page 36: ...34...
Page 74: ...72 20 EXTERNAL DIMENSIONS FZ 1200CT 700CT 500CT FX 1200CT 700CT 500CT Unit mm...
Page 80: ...78 MEMO...
Page 81: ...79 MEMO...
Page 82: ...80 MEMO...
Page 83: ......