![A&D FZ-1200CT Instruction Manual Download Page 73](http://html1.mh-extra.com/html/aandd/fz-1200ct/fz-1200ct_instruction-manual_3446893073.webp)
71
FX-1200CT
FX-700CT
FX-500CT
1260 ct
760 ct
510 ct
Weighing capacity
252 g
152 g
102 g
1260.042 ct
760.042 ct
510.042 ct
Maximum display
252.008 g
152.008 g
102.008 g
0.001 ct
Minimum weighing value (1 digit)
0.001 g
Repeatability (Standard deviation)
0.001 ct / 0.0005 g
Linearity
±0.002 ct / ±0.001 g
Stabilization time (typical at FAST)
*1
Approx. 2 seconds
Sensitivity drift
(10°C-30°C/50
°
F-86
°
F )
±2 ppm/°C
Internal mass
Unavailable
Clock function
Unavailable
Operating environment
5°C to 40°C (41°F to 104°F)
85%RH or less (No condensation)
Display refresh rate
*2
5 times/second or 10 times/second
Minimum unit mass
0.001 g
Counting
mode
*3
Number of samples
5, 10, 25, 50 or 100 pieces
Minimum 100% reference mass
0.100 g
Percent
mode
*3
Minimum 100% display
0.01%, 0.1%, 1% (Depends on the reference mass stored.)
External calibration weight
250 g 50 g
200 g 20 g
100 g
150 g 50 g
100 g 20 g
100 g 20 g
50 g
Weighing pan diameter
90 mm
Net weight
Approx. 3.4 kg
External dimensions
198 (W) x 294 (D) x 315 (H) mm
AC adapter
Confirm that the adapter type is correct for the local voltage and power
receptacle type
Power consumption
Approx. 11VA (supplied to the AC adapter)
Interface
RS-232C
*1
When MID., factory setting value of response rate, is used, the stabilization time is approx.
3.5 seconds.
*2
Factory setting: 5 times/second
*3
To use the counting mode or the percent mode, storing each mode as described in “9-9 Storing
Units” is required.
Summary of Contents for FZ-1200CT
Page 35: ...33 9 7 Data Format Examples LF...
Page 36: ...34...
Page 74: ...72 20 EXTERNAL DIMENSIONS FZ 1200CT 700CT 500CT FX 1200CT 700CT 500CT Unit mm...
Page 80: ...78 MEMO...
Page 81: ...79 MEMO...
Page 82: ...80 MEMO...
Page 83: ......