23
Operation
1 Press the ON:OFF key to turn off the display.
2 While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key.
The balance displays
p5
.
3 Press the PRINT key. Then the balance displays the function switches.
4 Set the switches using the following keys.
SAMPLE key
To select the switch to change the parameter.
RE-ZERO key
To change the parameter of the switch selected.
0
: To inhibit changes.
1
: To permit changes
PRINT key
To store the new parameter and return to the weighing mode.
CAL key
To cancel the operation (
Clr
is displayed.) To return to the
weighing mode, press the CAL key once again.
8-2 Initializing the Balance
This function returns the following parameters to factory settings.
Calibration data
Function table
The sample unit mass value (counting mode), 100% reference mass value (percent mode)
External calibration weight
Function switch settings
Statistical data
Internal mass correction value (Only for the FZ-CT series)
Note
Be sure to calibrate the balance after initialization.
Operation
1 Press the ON:OFF key to turn off the display.
2 While pressing and holding the PRINT key and the
SAMPLE key, press the ON:OFF key. The balance
displays
p5
.
3 Press the SAMPLE key to display
Clr
.
4 Press the PRINT key.
To cancel this operation, press the CAL key.
5 Press the RE-ZERO key.
6 Press the PRINT key to initialize the balance.
The balance automatically returns to the weighing
mode.